Illumination: 

Adjustable LED lighting integrated 

in base, universal 85 to 265 V, 50/60 Hz power 
supply 

Condenser:

 Abbe condenser N.A.1.25 NA 0.65 

with iris diaphragm , filter holder and blue filter, 
focussed via rack and pinion drive 

Dimensions: 

306 x 190 x 407 mm³ approx. 

Weight: 

6.6 kg approx.

 

 
 

3. Unpacking and assembly 

The microscope is packed in a molded styro-
foam container.  

 

Take the container out of the carton remove 
the tape and carefully lift the top half off the 
container. Be careful not to let the optical 
items (objectives and eyepieces) drop down.  

 

To avoid condensation on the optical com-
ponents, leave the microscope in the original 
packing to allow it to adjust to room tem-
perature.  

 

Using both hands (one around the pillar and 
one around the base), lift the microscope 
from the container and put it on a stable 
desk.  

 

The objectives will be found within individual 
protective vials. Install the objectives into the 
microscope nosepiece from the lowest 
magnification to the highest, in a clockwise 
direction from the rear.  

 

Put the head onto the top of the stand and 
tighten the head-lock-screw. Insert the eye-
pieces into the tube. 

 
 

4. Operation 

4.1 General information 

 

Set the microscope on a level table. 

 

Place the object to be observed in the centre 
of the specimen stage and clamp it to the 
object guide. 

 

Connect the mains cable to the net and turn 
on the switch to get the object illuminated. 

 

Make certain that the specimen is centered 
over the opening in the stage.  

 

Adjust the interpupillary distance so that one 
circle of light can be seen. 

 

Make the necessary eyepiece dioptre ad-
justments to suit your eyes. 

 

To obtain a high contrast, adjust the back-
ground illumination by means of the iris dia-
phragm and the variable illumination control. 

 

Rotate the nosepiece until the objective with 
the lowest magnification is pointed at the 

specimen. There is a definite “click” when 
each objective is lined up properly. 

NOTE: 

It is best to begin with the lowest power 

objective. This is important to reveal general 
structural details with the largest field of view 
first. Than you may increase the magnification 
as needed to reveal small details. 

When 100x 

(oil) objective is chosen, objective oil must be 
dripped onto the slide.

 

To determine the magnification at which you are 
viewing a specimen, multiply the power of the 
eyepiece by the power of the objective. 

 

Adjust the holding brake to give a suitable 
degree of tightness in the focusing mecha-
nism. 

 

Adjust the coarse-focusing-knob which 
moves the stage up until the specimen is fo-
cused. Be careful that the objective does not 
make contact with the slide at any time. This 
may cause damage to the objective and/or 
crack your slide. 

 

Adjust the fine-focusing-knob to get the im-
age more sharp and more clear. 

 

Colour filters may be inserted into the filter 
holder for definition of specimen parts. 
Swing the filter holder out and insert colour 
filters. 

 

Use the knobs of the mechanical stage to 
move the slide side-, back- and forwards. 
The vernier provides accurate location of the 
specimen area. 

 

Always turn off the light immediately after 
use. 

 

Be careful not to spill any liquids on the mi-
croscope. 

 

Do not mishandle or impose unnecessary 
force on the microscope. 

 

Do not wipe the optics with your hands. 

 

Do not attempt to service the microscope 
yourself. 

 

4.2  Using the polarisation equipment 

 

Insert the analyser into the slot on the re-
volving nosepiece. 

 

Place the polarising filter on the rim aperture 
of the light source. 

 

Rotate the polariser until the planes of the 
polariser and the analyser are exactly 
crossed, so that one sees a black back-
ground. 

Any object with a doubly-refracting (birefringent) 
structure should now appear brightly illuminated 
against the dark background. If that does not 
occur, it is possible that the direction of light 
vibration of the object coincides with the polari-
sation direction. Whether or not that is the case 

Summary of Contents for 500 1013146

Page 1: ...500 facher Vergrößerung Es ist mit einer Pola risationseinrichtung bestückt Stativ Robustes Ganzmetallstativ Stativarm fest mit Fuß verbunden Fokussierung über beidseitig am Stativ angebrachte koaxiale Stell knöpfe für Fein und Grobtrieb mit Kugellager und Feststellbremse einstellbarer Anschlag zum Schutz der Objektträger und Objektive Fokussierbereich 15 mm Einteilung der Feinfo kussierung 0 002 ...

Page 2: ... in den Strahlengang drehen Ein Klick Ton zeigt die richtige Stellung an Hinweis Es ist am besten mit der kleinsten Vergrößerung zu beginnen um zuerst größere Strukturdetails zu erkennen Der Übergang zu einer stärkeren Vergrößerung zur Betrachtung feinerer Details erfolgt durch Drehen des Revol vers bis zum gewünschten Objektiv Bei Ver wendung des Objektivs 100x muss Öl auf den Objektträger gegebe...

Page 3: ...rocke nen und staubfreien Platz aufbewahren Bei Nicht Benutzung das Mikroskop immer mit der Staubschutzhülle abdecken Das Mikroskop keinen Temperaturen unter 0 C und über 40 C sowie keiner relativen Luftfeuchtigkeit über 85 aussetzen Vor Pflege und Wartungsarbeiten ist immer der Netzstecker zu ziehen Zur Reinigung des Mikroskops keine ag gressiven Reiniger oder Lösungsmittel ver wenden Objektive u...

Page 4: ......

Page 5: ... polarisation equipment Stand Robust all metal stand with arm perma nently connected to the base Focussing by means of separate knobs for coarse and fine adjustment located on either side of the stand and operated by rack and pinion drive with ball bearings and retaining lever adjustable stopper for protecting the object slides and objective Focus range 15mm resolution of fine focussing adjustment...

Page 6: ...ecimen There is a definite click when each objective is lined up properly NOTE It is best to begin with the lowest power objective This is important to reveal general structural details with the largest field of view first Than you may increase the magnification as needed to reveal small details When 100x oil objective is chosen objective oil must be dripped onto the slide To determine the magnifi...

Page 7: ...holder in its socket 5 Storage cleaning and disposal Keep the microscope in a clean dry and dust free place When not in use always cover the micro scope with the dust cover Do not expose it to temperatures below 0 C and above 40 C and a max relative humid ity of over 85 Always unplug the mains plug before clean ing or maintenance Do not clean the unit with volatile solvents or abrasive cleaners Do...

Page 8: ......

Page 9: ...st équipé d un dispositif de pola risation Support Bâti support entièrement métallique Grande robustesse Mise au point rapide par pignon crémaillère à roulement à billes fine par vis micrométrique à commandes coaxiales de précision avec frein de serrage Butée de pro tection réglable Mise au point 15 mm gradua tion de la mise au point fine 0 002 mm Tube Tête de type Siedentopf binoculaire inclinée ...

Page 10: ... fort agrandissement pour voir des détails plus fins tourner le revolver jusqu à l objectif souhaité En cas d utilisation de l objectif 100x mettre de l huile sur le support de l objet La force d agrandissement résulte du produit du facteur d agrandissement de l oculaire et de l objectif Avec le frein réglez la tension appropriée du système de focalisation Le bouton de mise au point rapide permet ...

Page 11: ... 0 C et supérieures à 40 C ainsi qu à une humidité relative de l aide supérieure à 85 Avant d effectuer des travaux d entretien et de maintenance retirer toujours la fiche secteur Pour le nettoyage du microscope ne pas utiliser de nettoyants ni de solvants agressifs Pour le nettoyage ne pas démonter les objectifs ni les oculaires S il est fortement encrassé nettoyer le mi croscope avec un chiffon ...

Page 12: ......

Page 13: ...ione Stativo Robusto stativo completamente in me tallo braccio dello stativo fissato saldamente alla base messa a fuoco attraverso manopole coas siali per regolazione macrometrica e micrometri ca separate poste ai lati dello stativo con cusci netto a sfere e freno di arresto battuta regolabile per la protezione del portaoggetti e degli obiettivi Campo di messa fuoco 15 mm suddivisione della messa ...

Page 14: ...uminazione regolabile Ruotare l obiettivo con l ingrandimento mi nimo fino a portarlo sul percorso dei raggi luminosi Il raggiungimento della corretta posizione viene segnalato dallo scatto dell obiettivo Nota È opportuno cominciare con l ingrandimento minimo per poter riconoscere dapprima i dettagli macroscopici delle strutture Il passaggio a fattori di ingrandimento maggiori avviene attraverso l...

Page 15: ...ervazione pulizia smaltimento Conservare il microscopio in un luogo pulito asciutto e privo di polvere Durante il periodo di non utilizzo coprire se pre il microscopio con la custodia antipolvere Non esporre il microscopio a temperature inferiori a 0 C e superiori a 40 né ad un umidità relativa superiore all 85 Prima di effettuare lavori di cura o manuten zione è necessario staccare sempre la spin...

Page 16: ......

Page 17: ...veces Está dotado de un dispositivo de polarización Soporte Soporte de metal macizo brazo soporte de montaje fijo en el pedestal focalización por medio de ambos botones coaxiales para accionamiento burdo y fino con cojinete de bolas y freno de enclavamiento tope ajustable para protección del portaobjetos y de los objetivos Alcance de focalización 15 mm divisiones de la focalización 0 002 mm Tubo C...

Page 18: ...dio del diafragma de iris y la iluminación regulable Gire el objetivo con el menor factor de ampliación hacia el haz de luz Un sonido de clic indica la posición correcta Nota Es mejor empezar con la ampliación más pequeña para en primer lugar reconocer los detalles más burdos de la estructura El paso a una ampliación mayor para la observación de detalles más finos se efectúa girando el portaobjeti...

Page 19: ...sibles 5 Mantenimiento limpieza desecho El microscopio debe permanecer en un lugar limpio seco y libre de polvo Si se lo mantiene fuera de uso se debe cubrir siempre con la envoltura protectora contra el polvo No someta el microscopio a temperaturas bajo 0 C o sobre 40 C ni a una humedad relativa del aire superior al 85 Antes de realizar trabajos de mantenimiento o reparación se debe desconectar e...

Page 20: ......

Page 21: ...imais com um aumento de 40 até 1500 vezes É equipado com um dispositivo de polarização Tripé Pé de apoio inteiramente metálico e robusto braço de apoio fixado no pé ajuste focal por meio de botões de ajuste fino e grosseiro coaxiais a ambos lados do pé de apoio movidos por rolamentos e com freio de posicionamento limite de aproximação ajustável para a proteção das objetivas e dos suportes para obj...

Page 22: ...umento até a linha do raio luminoso O som de um clique indicará que a posição está correta Observação é melhor começar a observação com a objetiva de menor grau de ampliação de forma a reconhecer primeiro os detalhes estruturais maiores A passagem para um grau maior de aumento obtêm se girando o revólver até chegar na objetiva desejada Caso queira utilizar a objetiva de 100x deverá por óleo na pla...

Page 23: ...ira Quando não utilizado guarde sempre o microscópio na sua capa de proteção contra a poeira Não exponha o microscópio a temperaturas inferiores a 0 C ou superiores a 40 C assim como a uma humidade ambiente superior a 85 Antes de efetuar qualquer atividade de manutenção deve se sempre tirar o fio da tomada elétrica Não utilize produtos de limpeza agressivos ou solventes para limpar o microscópio N...

Page 24: ......

Reviews: