Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1
25 / 41
15.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Range
Description
Humidifier
Off, 1
~
5
There are five humidity levels available. As the
numbering increases, the humidity rises
accordingly. “Off” means the humidifier is
turned off. The default setting is “2”
Reslex
Off, 1
~
3
This feature enables the device to
automatically reduce the treatment pressure
when the patient exhales, so as to make the
user more comfortable. The higher the
numbering is, the more pressure the device
reduces. “Off” means this feature is disabled.
The default setting is “Off”
Ramp Time
0 - Max
Ramp
In order to increase comfort and help the
patient fall asleep easily, the pressure can
increase gradually, when the Ramp feature is
enabled. The ramp time during which the initial
pressure rises to the prescribed treatment
pressure can be adjusted. As you turn the Knob
to the nearest point, the numbering
increases or decreases by five minutes. The
default setting is “10 minutes.” The screen
displays a real-time countdown of the
remaining ramp time in minutes
Delay
On / Off
When the humidifier is on, this feature allows the
airflow to continue for about 15 minutes at a
low pressure (about 2 cmH
2
O) after you press
the Knob to discontinue treatment. This will
blow off the vapor left in the water chamber to
avoid any damage to the device. When this
feature is set to “Off,” which means it is
disabled, the airfolw stops delivering air instantly
after you press the Knob . The default setting
is “Off”
Date
2000-01-01
—
2099-12-31
Set date by adjusting this option
Time
00:00
—
23:59
Set time by adjusting this option
Brightness
High / Low
Setting screen brightness by adjusting this
option. The default setting is “High”
Mask Type
Full Face;
Nasal;
Pillow; Other
There are three mask types available, Full Face
(full-face mask), Nasal (nasal mask), and Pillow
(nasal pillow mask). The default mask type is
25
Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1