Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1
33 / 41
The patient is allergic to the
materials of the mask
Contact your physician and
equipment supplier.
Use a mask which is not made
with natural rubber latex.
Place a lining between the skin
and mask
Water in mask
When the humidifier is used,
the humidified air tends to
condense in the cold tube
and mask if the room
temperature is low
Turn the humidity setting down,
or raise the room temperature.
Place the tube under the quilt,
or use the tube cover.
Hang the tube loosely, and the
lowest part of the tube should
be lower than the patient's
head
Nasal, sinus, or
ear pain
Sinus or middle ear
inflammation
Contact your physician
immediately
Discomfort due
to inability to
adapt to the
treatment
pressure
The patient will feel
uncomfortable when the
treatment pressure is higher
than 13 cmH
2
O. However,
the treatment pressure is
determined according to
the patient's conditions,
and cannot treat sleep
apnea if the treatment
pressure is set too low
It takes a maximum of four
weeks to adapt to pressurized
air.
Relax and breathe through the
nose. If the problem still exists,
contact your physician
Obstructive
sleep apnea
symptoms recur
Probably because the
patient sleeps with his or
her mouth open, and the
pressurized air goes out via
the mouth, leading to
blockage in the respiratory
tract
Use a chin strap to prevent the
mouth from opening during
sleep, or use a full-face mask.
Contact your physician for
details
The device is
too noisy
The tube is not connected
properly
Reconnect the tube properly
Air delivered
from the device
is abnormally
hot
The air inlet of the device
may be partially blocked,
leading to insufficient
airflow into the device
Replace the air filter (see 15.6
Replacing the Air Filter), and
clean the air inlet
Place the device in an area
where air flows freely, and
make sure the device is at least
20 centimeters away from the
wall, curtain, or other things
33
Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1