7.1.2 Input / output signal cable
In order to ensure the sealing performance of the instrument access hole, the
single hole sealed cable should be a round multi strand sheathed cable with an outer
diameter of 6 ~ 10mm, and the double hole sealed cable should be a round multi
strand sheathed cable with an outer diameter of 3 ~ 5mm, with a copper core cross-
sectional area of 0.5 ~ 1.5mm2.
If ambient noise and crosstalk will adversely affect the signal, please use RVVP
shielded cable.
7.1.3
Temperature sensor signal cable
The signal cable of temperature sensor shall be a round multi strand sheathed
cable with an outer diameter of 3 ~ 5mm, and the cross-sectional area of copper core
is 0.2 ~ 0.3mm2.
7.1.4
Special signal cable for split flow sensor
five×0.5mm2rvvp special shielded cable, with an outer diameter of 8-10mm.
important