d.  0-2.5V (optional): The orange wire is the positive side of the 0-2.5V output and the white 

wire is the ground. See manual for setting the range on this output. 

 

3)  Attach the tubing connections: 

a.  Drain: Attach tubing from the drain outlet to waste. 
b.  Sample Inlet: Attach ozone compatible tubing to the sample inlet port. This tubing should 

be as short of a length as possible from your main system in order to reduce response 
time. 

c.  Return: If you are returning this to your system, you should use ozone compatible tubing 

and  make  sure  it  attaches  to  a  point  in  your  system  that  is  lower  pressure  than  where 
you attached the sample line. Otherwise, send to waste. 

d.  Air Inlet: If the instrument is being used in a hose-down environment, attach tubing and 

run it to an area that will not be exposed to water. Otherwise leave open. 

e.  Air exhaust: If the instrument is being used in a hose-down environment, attach tubing 

and run it to an area that will not be exposed to water. Otherwise, leave open. 

 

4)  Allow for warmup period: 

a.  Turn  the  instrument  on  and  allow  it  to  warmup  for  10-20  minutes  before  taking  ozone 

measurements.   