Model 106-L Ozone Monitor Manual Rev. E
13
To View the Flow Rate on the LCD
Rotate the scroll switch in either direction to view the volumetric flow rate during
normal operation. The flow rate will temporarily be displayed as follows:
O3= 5.3 ppb
Flow= 840 cc/min
To Transmit Logged Data to a Computer Using the USB or Serial Port
Connect the USB or serial port of the instrument to your computer using the
appropriate cable. Enable a data acquisition program on the computer such as
the 2B Technologies Display and Graphing Software, which can be
downloaded at:
http://twobtech.com/software.htm
Alternatively, HyperTerminal can be used (available on most Windows
platforms, usually in Start/All Programs/Accessories/Communications/Hyper
Terminal) or Terra Term Pro, which can be downloaded at:
http://hp.vector.co.jp/authors/VA002416/teraterm.html
The correct settings for receiving data are: chosen baud rate (2400, 4800 or
19200); 8 bits; no parity; 1 stop bit.
Click the Select button to obtain the
Main
Menu
. Go to the
Dat
submenu by
clicking on
Dat
. Next, click on
Xmt
. The message “Logged Data” will be
written to the serial port, followed by a carriage return and all of the lines of
logged data. After all data are transmitted, the message “End of Logged Data”
and a carriage return are written. After transmission is complete, you can
return to any position in the menu or resume ozone measurements. The
logged data continues to be available for transmission until a new data log is
started.
To Set the Calibration Parameters
The instrument is calibrated at the factory where slope (S) and offset (Z)
parameters are entered into the instrument’s memory. These preset calibration
parameters are given in the instrument’s Birth Certificate and recorded on the
calibration sticker on the back of the instrument. However, the calibration
parameters may be changed by the user. For example, it may be desirable to
provide a positive offset by a known amount (e.g., 10 ppb) if the analog output