ZyAura ZG1163R Скачать руководство пользователя страница 4

Ref.No.

012013

 

SPECIFICATIONS  

Method - Dual Beam NDIR (Non-Dispersive-Infrared) 
Display - LCD   

             

Independent CO

2, 

RH and Temperature readings. Calculates and Displays Ventilation Rates   

Sample Method - Diffusion or flow through (50 ~200 ml/min)   
CO

2

, Temperature & RH Specification: 

CO

2

 Specification

 

Measurement Range 

0-3,000 ppm display 

Display Resolution 

1ppm at 0~1,000ppm; 5ppm at 1,001~2,000ppm; 10ppm at 2,001~3,000ppm 

Accuracy 

0~2,000ppm: ±70 ppm or ±5% of reading, whichever is greater;

 

over 2000ppm: +/-7%   

Temperature Dependence 

Typ. ±0.2% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C 

Pressure Dependence 

0.13% of reading per mm Hg (Corrected via user input for altitude) 

Repeatability 

±20 ppm @ 400ppm 

Response Time 

About 2min for 63% of step change 

Warm-Up Time 

<60 seconds at 22°C 

Zone LED Display Green:<800ppm; 

Yellow:800~1200ppm;Red: 

>1200ppm 

Temperature Specification 
Temperature Range 

0°C to 50°C (32°F to 122°F)display  Display Resolution 0.1°C 

(0.1°F) 

 

 

Display Options °C/°F 

 

Response Time

 

20-30 minutes (case must equalize with environment)

 

Accuracy 

±1°C( ±2°F) When the fan blows to the device directly, the accuracy of temperature is + / -1.5 degC 

RH Specification 

Operating Conditions 

Measurement Range

 

20%-90% RH

 

Operating Temperature 

0°C to 50°C (32°F to 122°F) 

Display Resolution

 

1%RH

 

Humidity Range 

0 ~ 95% RH non-condensing 

Accuracy:

 

±5%RH@23°C 

Storage Conditions 

Response time:

 

<5 min for 63% of step change 

Storage Temperature 

-20°C to 60°C( -4°F to 140°F) 

Power Supply 

Power Supply 

100 ~ 240 VAC 50 / 60 Hz 
6 VDC from external AC/DC adapter which is included in package (Use specified AC adapter only)

 

Dimension 

 

 

Fault Codes & Troubleshooting Guide 

This section includes a list of Frequently Asked Questions for problems you may encounter with the ZG1163R CO2 Monitor. 

Fault Icon 

Description of the fault 

Suggested Actions 

 

The ambient temperature has exceeded the operating 
temperature range 0°C to 50°C (32°F to 122°F) 

This error will clear when the temperature returns to the range between 0°C to 
50°C (32°F to 122°F). 

 

 

EEPROM System Problem 

Please reconnect AC adapter to ZG1163R CO2 Monitor.   
If the “Err5, Err6” still appears, please contact the Service Department for further 
assistance. 

 
 

 

Radiant Innovation Inc.

  Http://www.ZyAura.com

 

1F, No.3, Industrial East 9

th

 Road, Science-Based Industrial Park, HsinChu, Taiwan 300.

 

Inch [mm] 