SPECIFICATIONS
Method - Dual Beam NDIR (Non-Dispersive-Infrared)
Display - LCD
Independent CO
2,
RH and Temperature readings. Calculates and Displays Ventilation Rates
Sample Method - Diffusion or flow through (50 ~200 ml/min)
CO
2
, Temperature & RH Specification:
CO
2
Specification
Measurement Range
0-3,000 ppm display
Display Resolution
1ppm at 0~1,000ppm; 5ppm at 1,001~2,000ppm; 10ppm at 2,001~3,000ppm
Accuracy
0~2,000ppm: ±70 ppm or ±5% of reading, whichever is greater;
over 2000ppm: +/-7%
Temperature Dependence
Typ. ±0.2% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C
Pressure Dependence
0.13% of reading per mm Hg (Corrected via user input for altitude)
Repeatability
±20 ppm @ 400ppm
Response Time
About 2min for 63% of step change
Warm-Up Time
<60 seconds at 22°C
Zone LED Display Green:<800ppm;
Yellow:800~1200ppm;Red:
>1200ppm
Temperature Specification
Temperature Range
0°C to 50°C (32°F to 122°F)display Display Resolution 0.1°C
(0.1°F)
Display Options °C/°F
Response Time
20-30 minutes (case must equalize with environment)
Accuracy
±1°C( ±2°F) When the fan blows to the device directly, the accuracy of temperature is + / -1.5 degC
RH Specification
Operating Conditions
Measurement Range
20%-90% RH
Operating Temperature
0°C to 50°C (32°F to 122°F)
Display Resolution
1%RH
Humidity Range
0 ~ 95% RH non-condensing
Accuracy:
±5%RH@23°C
Storage Conditions
Response time:
<5 min for 63% of step change
Storage Temperature
-20°C to 60°C( -4°F to 140°F)
Power Supply
Power Supply
100 ~ 240 VAC 50 / 60 Hz
6 VDC from external AC/DC adapter which is included in package (Use specified AC adapter only)
Dimension
Fault Codes & Troubleshooting Guide
This section includes a list of Frequently Asked Questions for problems you may encounter with the ZG1163R CO2 Monitor.
Fault Icon
Description of the fault
Suggested Actions
「
」
The ambient temperature has exceeded the operating
temperature range 0°C to 50°C (32°F to 122°F)
This error will clear when the temperature returns to the range between 0°C to
50°C (32°F to 122°F).
「
」
EEPROM System Problem
Please reconnect AC adapter to ZG1163R CO2 Monitor.
If the “Err5, Err6” still appears, please contact the Service Department for further
assistance.