- 17 -
Detailed accuracy indicator
Accuracy is specified for a period of one year after calibration, at 23
±
5
º
C, with relative humidity to 75 %.
Accuracy specifications are given as:
±
( [ % of reading ] + [ number of least significant digits ] ) (note:
“
number of
least significant digits
”
means the digits increased or decreased in least significant digits)
Measuring detailed accuracy indicator
Measuring
part
range
frequency
accuracy
remark
DC voltage
50V
10mV
0.5%+4
•
Measuring impedance 10 M
Ω
(standard
value)
•
Common mode restrain: 50Hz or 60Hz >
100Db
•
String mode restrain:50Hz or 60Hz >
45dB
•
Overload protection: 600V (peak
–
peak
value)
AC voltage
500V
100mV
0.5%+20 ( 45Hz~100Hz )
•
frequency beeps: 45Hz
~
400Hz
•
Adaptive to range from 10% to 100%
•
AC conversion: mean value
•
Measuring impedence:10 M
Ω
(standard
value)<100pF
•
Common mode restrain: 50Hz or 60Hz >
100Db
•
Overload protection: 600V(peak
–
peak
value)
2%+20 ( 100Hz~400Hz )