ZOLL M Series CCT
3
Using the M Series CCT options
The following sections describe how to use the M Series CCT option.
Changing the Display Traces
Trace 1 always displays the ECG waveform. You may select or change the waveform display for
Traces 2 and 3. If the unit has the appropriate options, you can choose from the following parameters:
• 3 Lead ECG (uses all three display channels)
• IBP Channel 1 (P1)
• IBP Channel 2 (P2)
• EtCO
2
• SpO
2
When the unit is in
PACER
or
DEFIB
mode, display messages take the place of Trace 3.
Any changes made remain in effect until either the settings are changed or for 10 seconds after the
M Series is turned off (the 10 second interval allows settings to remain in effect when the battery is
changed).
Trace 2 and 3 are factory configured to be OFF when the unit is powered up. However, they can be
configured to display specific parameters on power-up (refer to the
M Series CCT Configuration
Guide
).
To select the displayed waveforms:
1. Press the
Traces
softkey (if the “Traces” softkey label is not displayed, press the
Return
softkey until “Traces” displays, and press the
Traces
softkey).
The
Traces
menu display is shown in Fig. 3:
Figure 3: Traces Menu
2. Press the
Select
softkey to select “3 Lead ECG”, “Set Trace 2”, or “Set Trace 3.”
3. Press the
Enter
softkey.
4. If “3 Lead ECG” is selected:
• Traces 2 and 3 are set to ECG with each lead based on the active lead group. Trace 1 remains
set with the
LEAD
hard key. For example, if Trace 1 is set to Lead II with the
LEAD
hard
key, Trace 2 and 3 to are set to Lead I and Lead III respectively.
• For units with 12 Lead ECG option, Custom Lead groups may be set based on preconfigured
choices. For additional information, refer to the
M Series CCT Configuration Guide
and
12
Lead Operator’s Guide
.