ZOLL M Series CCT

Recommended Separation Distances from RF Equipment for 

the M Series CCT Life-Support Functions

Recommended separation distances between portable and mobile RF communications equipment 
and the M Series CCT.

The life-support functions

a

 of the M Series CCT are intended for use in the electromagnetic environment in 

which radiated RF disturbances are controlled. The customer or user of the M Series CCT can help prevent 
electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the M Series CCT as recommended below, according to the 
maximum output power of the communications equipment.

a. The life-support functions on the M Series CCT are defined to be any function associated with ECG monitoring,
pacing, defibrillation, and shock analysis. Specifically, these functions include, but are not limited to, the ECG
waveform monitoring from leads or pads, the pacing pulse output, QRS detection, defibrillation energy discharge, and
shock advisory functions.

Rated maximum 

output power of 

equipment

(in watts)

Separation distance according to frequency of transmitter

(in meters)

150 kHz to 80 MHz

outside ISM bands

150 kHz to 80 MHz

in ISM bands

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.035

0.12

0.06

0.12

0.1

0.11

0.38

0.19

0.38

1

0.35

1.2

0.6

1.2

10

1.1

3.8

1.9

3.8

100

3.5

12

6

12

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in 

meters can be determined using the equation applicable to the frequency of the transmitter, where 

P

 is the 

maximum output power rating of the transmitter in watts according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
NOTE 3: An additional factor of 10/3 is used in calculating the recommended separation distances for 
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 
2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if 
it is inadvertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects, and people.

d

 =0.35 

P

 

d

 =1.2 

P

 

d

 =0.6 

P

 

d

 =1.2 

P

Содержание M Series CCT

Страница 1: ...M Series CCT ZOLL M Series CCT 9650 0222 01 Rev D...

Страница 2: ...date or revision level for this operation guide is shown on the front cover ZOLL and M Series are registered trademarks of ZOLL Medical Corporation Copyright 2007 ZOLL Medical Corporation All rights...

Страница 3: ...to an External VGA Display monitor 5 Configuring Displayed Parameters by Color 6 MAINTENANCE 7 SPECIFICATIONS 7 IEC 60601 1 2 Specifications 9 Electromagnetic Emissions Declaration 9 Electromagnetic...

Страница 4: ...iv ZOLL M Series CCT...

Страница 5: ...sonnel only Do not use in the presence of oxygen rich atmospheres flammable anesthetics or other flammable agents such as gasoline Do not use near the site of a gasoline spill Explosion may result Avo...

Страница 6: ...CCT in an aircraft fluctuations in cabin pressure can affect invasive blood pressure IBP calibration When cabin pressure fluctuates significantly due to altitude changes you must rezero the IBP trans...

Страница 7: ...the M Series Operator s Guide for the controls available on both the M Series and the M Series CCT M Series CCT Option In addition to the standard M Series features the M Series CCT option can do the...

Страница 8: ...on the display The M Series CCT three Trace display areas are shown in Fig 2 The trace type for each display is as follows Trace 1 reserved for ECG Trace 2 type of trace determined by user Trace 3 ty...

Страница 9: ...when the unit is powered up However they can be configured to display specific parameters on power up refer to the M Series CCT Configuration Guide To select the displayed waveforms 1 Press the Trace...

Страница 10: ...2 or Set Trace3 is selected The Trace Options menu displays see Fig 4 Figure 4 Trace Options Menu 6 Press the Select softkey to select the type of waveform you want to display You cannot select a wave...

Страница 11: ...Series CCT is initially powered up To enable or disable the VGA output port 1 Press the CONTRAST button see Fig 1 on page 1 the Contrast menu appears see Fig 5 Figure 5 Contrast menu 2 To enable activ...

Страница 12: ...asive pressure 1 and pressure 2 Temperature etc The parameter display colors available are Red Cyan Purple Blue Green Yellow and White Parameter display colors are factory configured as indicated belo...

Страница 13: ...powered per EN 60601 1 Class II and internally powered per EN 60601 1 DC input only Design Standards Meets or exceeds UL 2601 AAMI DF 39 AAMI DF 2 IEC 601 2 4 EN 60601 2 25 and EN 60601 2 27 Patient S...

Страница 14: ...and immediately put into use Humidity 5 to 95 relative humidity noncondensing Vibration Mil Std 810 E Minimum Integrity Test Shock IEC 68 2 27 50g 6mS half sine Operating Pressure 594 to 1060 milliba...

Страница 15: ...ronment Emissions test Compliance Electromagnetic environment guidance RF Emissions CISPR 11 Group 1 The M Series CCT uses RF energy for its internal function only Therefore its RF emissions are very...

Страница 16: ...1000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupt...

Страница 17: ...tter in watts according to the transmitter manufacturer and d is the recommended separation distance in meters c Field strengths from fixed RF transmitters as determined by electromagnetic site survey...

Страница 18: ...rs 150 kHz to 80 MHz outside ISM bands 150 kHz to 80 MHz in ISM bands 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 035 0 12 0 06 0 12 0 1 0 11 0 38 0 19 0 38 1 0 35 1 2 0 6 1 2 10 1 1 3 8 1 9 3 8 100 3...

Страница 19: ...Hz d 0 47 800 MHz to 2 6 GHz where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distance in meters Fie...

Страница 20: ...t Functions table Note a Specifically these functions are the Invasive Blood Pressure IBP channels the Temperature channels the Non invasive Blood Pressure NIBP End Tidal CO2 EtCO2 and SpO2 Rated maxi...

Страница 21: ...ic compatibility EMC specification 8 discharge specification 8 emissions declaration 9 immunity declaration 10 immunity specification 8 Emissions protection radio frequency v Enable activate VGA outpu...

Страница 22: ...ace display area description of 2 illustration 2 Traces illustration of menu 3 softkey 3 to display menu 3 U Usage time XL Battery 7 V VGA connecting display 5 port enable time out during power loss 5...

Отзывы: