On M Series CCT units, you can set the NIBP cuff inflation pressure for three patient type
modes: adult, pediatric and neonatal. With NIBP First Inflate Pressure Default highlighted and
Set/Review
selected, press the
Change Value
softkey. The following screen displays:
Pressing the
Change Value
softkey will scroll through the available inflation pressures for the
highlighted parameter.
Pressing the
Prev Item
or
Next Item
softkeys moves the highlight to the previous or next
parameter in the menu.
Pressing the
Return
softkey selects the values that have been entered and returns to the main
configuration menu.
Basic Auto Energy Escalation
When set to Yes, the unit automatically increments the defibrillation energy to the levels
specified in Energy Level: Shock 1, 2, 3 after each of the first two shocks and displays the
message “ENERGY INCREMENTED,” when both of the following are true:
•
The defibrillator is in Manual mode
•
The defibrillator has external paddles or pads attached
Manually changing the energy level outside the pre-programmed sequence and delivering a
shock disables this function until the unit is turned off for more than 10 seconds, then turned
back on.
This option increments the energy level regardless of Shock Advisory activation or status. It
does not work with internal handles.
For more information about the energy level settings, see the “Energy Level: Shock 1, 2, 3”
section of this manual.