M S
ERIES
CCT C
ONFIGURATION
G
UIDE
8
Set Report Restart Delays
Allows review of, or changes to the length of time the device must be turned off before it erases
all data stored in either summary report memory or trend memory.
With Set Report Restart Delay highlighted and Set/Review selected, press the
Change Value
softkey. The following screen displays:
Pressing the
Change Value
softkey will scroll the highlight through the available delay times.
Pressing the
Prev Item
and
Next Item
softkeys will scroll the highlight to the next or previous
restart delay feature.
Pressing the
Enter
softkey will cause the unit to return to the main configuration menu.
Print 12 Lead 4x3
Determines how many copies of a 12-lead 4x3 report will print after pressing and holding the
RECORDER
button for three seconds. When set to “1 Copy” or “2 Copies” one or two copies
of the 12-lead 4x3 report will print. When set to “None”, pressing and holding the
RECORDER
button will cause a one or three lead DIAG BANDWIDTH ECG strip to be
printed.
Auto Self-Test Timeout Period
Allows the user to set the time between tests or disable the automatic tests. The unit is capable
of automatically running a self-test at seven to thirty day intervals. If the unit is off, plugged in,
and the test connector is attached to the Multi-Function cable, the unit will test the battery
status, charge and discharge the defibrillator, measure the amplitude and timing of the pacing
waveform, and measure the impedance of the test port.
Note:
This option is not yet implemented.
Async Softkey in Pace Mode
Sets the unit to display the Async Pacing On/Off softkey when the unit is in pace mode.
Code Markers
Allows review of or changes to the preconfigured list of Code Markers available for use in
PACER, MONITOR, and DEFIB modes.
Note:
MONITOR Code Markers are not used in AED Models.