ZOLL M Series Invasive Blood Pressure
Specifications
For a list of alarm defaults and ranges, see "Setting Alarms" on page 11.
Table 5: General Specifications
Principle of Operation
Resistive Bridge transducer; 5uV/V/mmHg sensitivity
(transducers with 40uV/V/mmHg sensitivity are
not
compatible)
Regulatory Standards
EN60601-2-34:1995
Table 6: IBP Measurement Specifications
Pressure Range
-50 to 300 mmHg after offset adjustment
Zero Offset Adjust Range
+/- 200 mmHg
Accuracy
Greater of +/- 2% measurement or +/- 2 mmHg, plus transducer
error
Baseline Stability
+/- 0.1 mmHg/
°
C excluding transducer drift
Bandwidth
DC to 20 Hz
Noise
< 1 mmHg peak-to-peak
Transducer Impedances
Input: minimum 250
Ω
Output: maximum 3000
Ω
Transducer Excitation
2.5 V DC
Table 7: Battery Operating Time Specifications (using ZOLL XL Battery Pack)
For a new, fully charged battery
pack at 20°C with IBP, EtCO
2
,
SpO
2,
NIBP and Temperature
options
60 defibrillator discharges at maximum energy (200J Biphasic),
or 1.75 hours minimum of continuous ECG, IBP (2 channels),
EtCO
2
, SpO
2
, and Temperature monitoring, or 2.00 hours of
continuous ECG monitoring/pacing at 60 mA, 70 beats/min.
Table 8: Environmental Specifications
Operating Temperature
0
°
C to 50
°
C
Storage and Shipping Temperature
-20
°
C to 60
°
C
Electromagnetic Immunity
AAMI DF-2: 1996, EN 61000-4-3: 1996, 15 V/m
Humidity
5% to 95% relative humidity, non-condensing
Operating Pressure
594 mBar to 1060 mBar