ICY

® 

Intravascular Heat Exchange Catheter (Custom Luer) 

Instructions for Use 

Model IC-3893A/8700-0782-01 

 

10 of 10 

 

 

 

 

ICY 

 

 

 

SilvaSorb 

 

Содержание ICY IC-3893A

Страница 1: ...t the sterility barrier has not been compromised Storage Store between 20 C 25 C Avoid freezing and excessive heat above 40 C Indications for Use The ZOLL ICY Intravascular Heat Exchange Catheter Model IC 3893A connected to the ZOLL Coolgard 3000 Thermogard XP Thermal Regulation System is indicated for use In cardiac surgery adult patients to achieve and or maintain normothermia during surgery and...

Страница 2: ...ive force in moving the catheter or guidewire If resistance is encountered an x ray should be performed to identify the reason for the resistance 10 Passage of the guidewire into the right heart can cause dysrhythmias right bundle branch block vessel wall atrial or ventricular perforation 11 Use only sterile normal saline for catheter priming It is the circulating fluid in the catheter 12 The cath...

Страница 3: ... connect to standard Luer Lock syringes or other standard Luer Lock connectors 4 Remove the caps from the IN and OUT Luers With the catheter cover in place fill the syringe 5 cc or larger with sterile saline and attach the syringe to the female IN Luer 5 WARNING Never inject positive pressure into the IN Luer with the OUT Luer cap in place 6 Gently inject saline through the catheter until it begin...

Страница 4: ...indwelling catheter length using the centimeter marks on the catheter shaft as reference Frequent visual reassessment should be made to ensure that the catheter has not moved 26 Attach a primed Start Up Kit to the catheter by connecting the male Luer of the Start Up Kit to the female IN Luer of the catheter and the female Luer of the Start Up Kit to the male OUT Luer of the the catheter White ZOLL...

Страница 5: ...t move the catheter if resistance is felt Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped If they are capped uncap them deflate the balloon and try removing the catheter again If resistance is still encountered an x ray should be performed to identify the reason for the resistance MRI Safety Information Non clinical testing has demonstrated that the ICY catheter is ...

Страница 6: ...e guidewire at all times Dispenser Every guidewire is provided in a dispenser package Remove the guidewire anti migration clip before dispensing the guidewire Remove the guidewire protective cap immediately prior to guidewire use Prepare the guidewire prior to insertion It is recommended that the dispenser be filled with heparinized solutions e g saline or dextrose to bathe the guidewire during in...

Страница 7: ...xchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 7 of 10 1 OUT 2 IN 3 Medial 4 Distal 5 Proximal 6 Detail view ICY IC 3893A Catheter Shaft 7 Balloon 5 mm x 80 5 mm 8 Balloon 8 mm x 80 5 mm Balloons ...

Страница 8: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 8 of 10 ...

Страница 9: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 9 of 10 ...

Страница 10: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 10 of 10 ICY SilvaSorb ...