Monitoring the ECG Rhythm
9650-0350-01 Rev. J
ZOLL AED Pro Operator’s Guide
5–7
Monitoring with Defibrillation Electrodes
With defibrillation electrodes in use, you can change from semiautomatic mode to ECG
monitoring mode.
If the unit detects a shockable rhythm while in ECG monitoring mode and defibrillation
electrodes are in use, the unit prompts you to check the patient and then changes to
semiautomatic mode.
For more information, refer to Chapter 3, “Semiautomatic Mode”.
Before you begin
Attach defibrillation electrode pads to the patient (refer to Chapter 3.)
Procedure
To change from semiautomatic to ECG monitoring mode:
Step
Action
Result
1
Press and hold the left (unlabeled) softkey
for at least 5 seconds.
The unit displays
MONITOR
as the current
mode, and the
SEMI-AUTO
softkey label
appears.
2
Monitor the patient’s ECG.
If it detects a shockable rhythm while
monitoring, the unit prompts you to check the
patient and changes to semiautomatic mode.
3
To manually return to semiautomatic mode,
press and release the
SEMI-AUTO
softkey.
00:24:18
1
34
MONITOR
SEMI-AUTO
ECG X1.0
CHECK PATIENT