Before you begin
Be sure the defibrillation cable is plugged into the unit.
Procedure
To defibrillate the patient in semiautomatic mode:
Step
Action
Result
1
Press and release the
On/Off
button to turn
on the unit.
On successful completion of the power-on
self-test, the unit indicates the type of
attached electrodes. If the defibrillation
pads are not attached to the patient, the
unit issues the voice and text prompt:
ATTACH DEFIB PADS TO PATIENT’S
BARE CHEST
2
If prompted, apply defibrillation electrode
pads to the patient.
(Refer to “Applying Defibrillation Electrode
Pads” on page 3-2.)
When the pads are properly attached to the
patient, the unit issues the voice and text
message:
DON’T TOUCH PATIENT, ANALYZING
and then begins ECG analysis.
3
Allow the unit to analyze the patient’s ECG.
The unit determines whether or not the
patient has a shockable rhythm and then
displays and voices its recommendation:
SHOCK ADVISED
or
NO SHOCK ADVISED
4
After ECG analysis, which message
appears?
— If SHOCK ADVISED, continue with step 5.
— If NO SHOCK ADVISED, go to step 8.
If a shock is needed, the defibrillator
automatically begins charging.
If a shock is not needed, the unit prompts
you to perform CPR.
5
Wait for the defibrillator to charge.
When fully charged, the unit emits a
charge-ready tone, repeatedly flashes the
Shock
button, and issues the voice and
text prompts:
DON’T TOUCH PATIENT
PRESS FLASHING SHOCK BUTTON
Содержание aed pro
Страница 1: ...AED Pro AED Pro A W Operator s Guide 9650 0350 01 Rev J...
Страница 6: ...TABLE OF CONTENTS iv www zoll com 9650 0350 01 Rev J...
Страница 54: ...CHAPTER 3 SEMIAUTOMATIC MODE 3 14 www zoll com 9650 0350 01 Rev J...
Страница 118: ...AED PRO OPERATOR S GUIDE 9650 0350 01 Rev J ZOLL AED Pro Operator s Guide C 10 This page intentionally left blank...