background image

9650-000761-01 Rev. A

ZOLL AED 3 Service Manual

5

Cleaning the AED

After each use, clean and disinfect the defibrillator with a soft, damp cloth using 
either 90% isopropyl alcohol, or soap and water. You can also use a chlorine 
bleach and water mixture (30 ml/liter water) to clean the defibrillator (except on 
the contacts and connectors).

NOTE Wipe down the defibrillator with water after using any cleaning solution. 

Chlorine residue that remains on the LCD screen may cause damage.

DO NOT:
• Immerse any part of the defibrillator in water.
• Use chlorine mixture on contacts or connectors; this will degrade the contacts 

over time.

• Use ketones (MEK, acetone, etc.) to clean the defibrillator.
• Use abrasives (e.g., paper towel) on the display window or graphic screen.
• Sterilize the defibrillator.

Table 1. Self Test Functions

Battery 
Installation 
Self Test

Power On 
Self Test

Manual 
Self Test

Automatic 
Self Test

Automatic 
Monthly 
Test

Battery Capacity

Defibrillation Pads
Connection

Battery Expiration

ECG Circuitry

Defibrillator Charge 
and Discharge Circuitry
(2 Joule Charge/
Discharge Test)

Microprocessor
Hardware/Software

CPR Circuitry and
Sensor (if defibrillation 
pads with CPR func-
tionality are connected)

Audio Circuitry

200 Joule Charge/
Discharge Test

Содержание AED 3

Страница 1: ...BLS Service Manual 9650 000761 01 Rev A...

Страница 2: ...R D padz Pedi padz PlusTrac Program Management Onboard RescueNet Stat padz ZOLL and ZOLL AED 3 are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other cou...

Страница 3: ...ment Needed 8 Set Up 8 Pediatric Mode Test 10 Equipment Needed 10 Set Up 10 Energy Test 11 Equipment Needed 11 Set Up 11 Wi Fi Test 14 Equipment Needed 14 Set Up 14 USB Test 16 Audio Recording Test ZO...

Страница 4: ...ev A Safety Function 23 Safety on High Voltage Capacitor Charge 23 Safety on Discharge 24 Safety on Digital Circuits 24 Power Supply 25 Appendix Hardware System Architecture 27 Overview 27 ZOLL AED 3...

Страница 5: ...to its automatic self tests However this manual provides optional maintenance tests you can perform periodically according to local protocols This manual contains the following sections Preface Conta...

Страница 6: ...efore using this defibrillator read the ZOLL AED 3 Operator s Manual Before servicing any equipment review these safety considerations and read this manual carefully Before discharging the defibrillat...

Страница 7: ...structions are included in this chapter Self Tests Cleaning Manual Maintenance Tests Self Tests The ZOLL AED 3 defibrillator performs the following self tests to verify AED integrity and its readiness...

Страница 8: ...electronics are functional Defibrillator Charge and Discharge Circuitry Verifies that the device s defibrillator electronics are functional and can charge and discharge at 2 joules The following test...

Страница 9: ...tor in water Use chlorine mixture on contacts or connectors this will degrade the contacts over time Use ketones MEK acetone etc to clean the defibrillator Use abrasives e g paper towel on the display...

Страница 10: ...o o 3 Are input connectors clean and undamaged o o 4 Are there any cracks or loose parts in the housing o o 5 Does the front panel display have any damage or cracks o o ELECTRODES 6 Is the cable free...

Страница 11: ...maintenance tests Fluke Impulse 7000 Defib Analyzer or equivalent Stopwatch ZOLL AED 3 Defib Test Cable Kit REF 8900 000268 Blank USB Flash Drive minimum 512 MB USB 1 1 or 2 0 compliant Computer Wi Fi...

Страница 12: ...ture 4 Set the Power Supply to 12 0V then turn it off 5 Connect the leads from the Battery Test Fixture to the proper polarity of the Power Supply and install the Battery Test Fixture into the battery...

Страница 13: ...the AED powers on in Rescue Mode Note The AED may issue the CHANGE BATTERY message Disregard the message and continue the test o o 6 On the Battery Test Fixture turn the I Measure switch to the sleep...

Страница 14: ...to the AED then press the On Off button to turn on the AED Verify that the AED issues the prompts UNIT OK ADULT PATIENT SELECTED followed by IF PATIENT IS A CHILD PRESS CHILD BUTTON o o 2 Press the Ch...

Страница 15: ...at the AED issues the prompts UNIT OK and PLUG IN PADS CABLE o o 2 Connect the ZOLL AED 3 Defib Test Cable Kit to the AED Verify that the AED issues the voice prompt ADULT PATIENT SELECTED followed by...

Страница 16: ...e energy delivered into analyzer is 123 joules to 167 joules o o 12 Verify that the AED issues the prompt START CPR after 2 minutes default setting verify that the AED issues the prompt STOP CPR o o 1...

Страница 17: ...Selected Energy 50 J 70 J 85 J 120 J 150 J 200 J 25 37 J 54 J 57 J 86 J 109 J 139 J 50 50 J 69 J 80 J 118 J 145 J 209 J 75 61 J 82 J 97 J 134 J 166 J 196 J 100 60 J 84 J 95 J 142 J 165 J 194 J 125 57...

Страница 18: ...the AED displays the Configuration window o o 3 Press the Supervisor icon and enter the six digit passcode default passcode is 123456 and press OK Verify that the AED displays the Advanced Configurati...

Страница 19: ...Verify that the AED displays the TESTING SERVER CONNECTIVITY message along with the IN PROGRESS message o o 10 Verify that the AED displays the message TESTING SERVICE CONNECTIVITY COMPLETE o o 11 Pre...

Страница 20: ...o 2 Press the Export Files icon and select Clinical Archives all Verify that the green check appears next to the Clinical Archives all option o o 3 Press Save USB When the transfer is complete verify...

Страница 21: ...fibrillator with the Audio Recording option is capable of recording up to 120 minutes of continuous ECG Audio and clinical event data during the rescue of a patient case Action Expected Result s Pass...

Страница 22: ...18 ZOLL AED 3 Service Manual 9650 000761 01 Rev A...

Страница 23: ...nder Operator Action If these steps do not remedy the problem follow the suggestions under Technical Action Table 3 ZOLL AED 3 Troubleshooting Problem Operator Action Technical Action Green check does...

Страница 24: ...orting the patient by stretcher or vehicle stop all patient movement during analysis If the message persists remove the AED from service NO SHOCK DELIVERED The Shock button was not pressed or an inter...

Страница 25: ...controller MCU and the Complex Programmable Logic Device CPLD in real time The MCU gets instructions commands from the System on Module SoM such as charge voltage for the high voltage capacitor discha...

Страница 26: ...side signals CHARGE XEND_CYCLE CHARGE_CURRENT Power supply non isolated from battery pack Circuits on high voltage side also referred to as isolated side are MCU CPLD ECG circuitry acquiring ECG data...

Страница 27: ...independent circuits monitor the high voltage capacitor voltage Vcap1 Vcap2 This arrangement provides redundant measurement of high voltage capacitor voltage If the measurements of Vcap1 and Vcap2 are...

Страница 28: ...ate is also monitored by GPIO channels in the MCU and or CPLD Any single failure does not cause the patient relay to accidentally be turned on and thus prevents the patient from receiving an inappropr...

Страница 29: ...Power Supply There are two power supplies one is for the SoM non isolated from battery pack and the other one for the ECG CPLD MCU components isolated from battery pack Power consumption in the monito...

Страница 30: ...26 9650 00761 01 Rev A...

Страница 31: ...9650 000761 01 Rev A ZOLL AED 3 Service Manual 27 Appendix Hardware System Architecture Overview This appendix includes the Hardware System Architecture diagram...

Страница 32: ...citor Patient Relay SOM i MX28 with WiFi Cable ID Discrete ECG Patient Impedance and Pace Detect USB Device USB Host RBG I2C Antenna GPIO Readiness Indicator BAT BAT I2C SMBUS Bus I2C Defib Discharger...

Страница 33: ...ance tests and keep it for your records Result of Check o No action required o Minor problems corrected o Disposable supplies replaced o Major problems identified unit out of service Additional Remark...

Страница 34: ...o 10 o o 11 o o 12 o o Pass Fail N A 1 o o o 4 o o o 5 o o o 6 o o o 7 o o o Pass Fail 1 o o 2 o o 3 o o 4 o o 5 o o 6 o o 7 o o Pass Fail 1 o o 2 o o 3 o o 4 o o 5 o o 6 o o 7 o o 8 o o 9 o o 10 o o...

Отзывы: