9650-000761-01 Rev. A
ZOLL AED 3 Service Manual
5
Cleaning the AED
After each use, clean and disinfect the defibrillator with a soft, damp cloth using
either 90% isopropyl alcohol, or soap and water. You can also use a chlorine
bleach and water mixture (30 ml/liter water) to clean the defibrillator (except on
the contacts and connectors).
NOTE Wipe down the defibrillator with water after using any cleaning solution.
Chlorine residue that remains on the LCD screen may cause damage.
DO NOT:
• Immerse any part of the defibrillator in water.
• Use chlorine mixture on contacts or connectors; this will degrade the contacts
over time.
• Use ketones (MEK, acetone, etc.) to clean the defibrillator.
• Use abrasives (e.g., paper towel) on the display window or graphic screen.
• Sterilize the defibrillator.
Table 1. Self Test Functions
Battery
Installation
Self Test
Power On
Self Test
Manual
Self Test
Automatic
Self Test
Automatic
Monthly
Test
Battery Capacity
Defibrillation Pads
Connection
Battery Expiration
ECG Circuitry
Defibrillator Charge
and Discharge Circuitry
(2 Joule Charge/
Discharge Test)
Microprocessor
Hardware/Software
CPR Circuitry and
Sensor (if defibrillation
pads with CPR func-
tionality are connected)
Audio Circuitry
200 Joule Charge/
Discharge Test