9650-000761-01 Rev. A
ZOLL AED 3 Service Manual
17
Audio Recording Test
If the defibrillator does not have audio recording functionality enabled, record
N/A for this test on the Maintenance checklist on the back page of this
manual.
Equipment Needed
USB Flash drive (containing Clinical Event data from the previous USB test),
Power Supply, Battery Test Fixture, PC (with RescueNet Code Review
version =>5.72).
Set Up
Audio recording must be set to ON; the default for this configuration setting is
ON. Refer to the Configuration Settings in the ZOLL AED 3 Administrator’s Guide
if you need to change this setting to ON.
NOTE
The ZOLL AED 3 BLS defibrillator (with the Audio Recording option) is
capable of recording up to 120 minutes of continuous ECG/Audio and
clinical event data during the rescue of a patient case.
Action
Expected Result(s)
Pass/Fail
1.
Insert USB Flash Drive (containing
the Clinical Event data from the
previous USB test) in the
computer USB port. Open File
Explorer, and select the USB
drive.
Verify that CRD files are present.
o
o
2.
Double-click the desired file and
then click Yes in the message
box.
Verify that the RescueNet Code
Review Standard Edition window
appears.
o
o
3.
Select 3-Magnified ECG at the top
of the screen and click Play.
Verify that you can hear all voice
prompts and events (from the
computer speaker) that were
recorded by the AED.
o
o
4.
Press the On/Off button to turn off
the AED.
Verify that the AED displays the
message UNIT POWERING OFF.
o
o