Self Tests
The ZOLL AED 3 defibrillator performs the following self tests to verify
unit integrity and its readiness for emergency use. See the
ZOLL AED 3
Administrator’s Guide
for more detailed information about these tests.
Table 2: ZOLL AED 3 Self Tests
Self Test
When Performed
Functions Performed
Battery Installation
Automatically, after battery
pack is installed or
replaced.
Basic functions:
• Battery Capacity
• Defibrillation Pads
Connection
• Defibrillation Pads/
Battery Expiration
Verification
• ECG Circuitry
• Defibrillator Charge and
Discharge Circuitry
(2 Joule Test)
• Microprocessor
Hardware/Software
• CPR Circuitry and Sensor
(if defibrillation pads with
CPR functionality are
connected)
• Audio Circuitry
Additional function:
• 200 Joule Charge Test
Power On
Automatically, whenever
the AED is turned on.
Basic functions:
Same as Battery Installation
Self Test basic functions.
Manual
By pressing and holding the
AED’s On/Off button for at
least 5 seconds.
Basic functions:
Same as Battery Installation
Self Test basic functions.
Additional function:
• 200 Joule Charge Test