![Zolix Instruments Omni-l200i Series Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/zolix-instruments/omni-l200i-series/omni-l200i-series_operation-manual_3986095011.webp)
Installation and Operation
8
Zolix Instruments Co., Ltd.
http://www.zolix.com.cn E-mail: [email protected]
Slit house
The width of the slit is one of the most important parameters for resolution of the monochromator. The
slit width of each bilateral slit assembly is adjustable from 0.010 millimeters to 3 millimeters (10µm to
3,000µm) by a micrometer located on the top of the slit housing. For maximum reproducibility, the slit
width should be set in a clockwise direction (increasing slit widths) each time it is changed.
Note: Damage may be done if slit jaws are opened wider than 3.0mm.
Содержание Omni-l200i Series
Страница 14: ...Installation and Operation 11 Zolix Instruments Co Ltd http www zolix com cn E mail info zolix com cn...
Страница 16: ...Installation and Operation 13 Zolix Instruments Co Ltd http www zolix com cn E mail info zolix com cn...
Страница 20: ...Dimensions 17 Zolix Instruments Co Ltd http www zolix com cn E mail info zolix com cn 8 Dimensions...