Installation and Operation
7
Zolix Instruments Co., Ltd.
http://www.zolix.com.cn E-mail: [email protected]
5.1.2 Connecting the spectrograph
a. Firstly ensure the Power switch on the rear of the spectrograph is set to 0
b. Connect the Power & USB cables.Connect the other end of the USB cable to the PC and plug in
the power supply to the mains, switch the power on and install the USB driver as USB SOFTWARE
INSTALLATION.
c.
Remove the input slit and output slit covers from the spectrograph, e.g.:
d. Fitting the light source to the input slit. Then the system will be as Tunable Light Source. We can
supply Deuterium, Multiplex,Tungsten-Halogen and Xenon light source.
e. Fitting the detector to the output slit, we can supply Photomultiplier,Silicon Series
Detectors,InGaAs Series Detectors and so on.
f.
Open the light source, spectrometer, data collector power, open the CTL-C1 or Zolixscan
scanning software , you can begin to use.
5.2. Accessory Operation
5.2.1 Slit
Micrometer
The micrometer knob is graduated in 0.010-millimeter (10µm) increments. Each clockwise revolution
of the micrometer knob increases the slit width 0.25 millimeters (250µm). Refer to the side drawing.
The micrometer knob should not be rotated below a reading of 0.00 or above 3.00. A micrometer
setting of less than 0.010 millimeters (10µm) should not be used, because a stop is provided to
prevent the slit jaws from contacting each other.