Error Codes

Display message What it means

What to do

• The sample volume is too small
•  There may be a problem with the 

test strip

• Meter error

•  Use a new test strip and try again with proper amount of 

sample

•  Perform a control solution test using a new test strip. If the 

test results are within the range printed on the test strip vial, 
retest using a fresh blood sample and a new test strip

•  If the control solution result is out of range or the error 

reappears, contact Zoetis Customer Care

• Meter error
•  Internal reference error 

(data memory)

•  Firmware checksum error 

(program memory)

•  Perform a control solution test using a new test strip. If the 

test results are within the range printed on the test strip vial, 
retest using blood and a new test strip

•  If the control solution result is out of range or the error 

reappears, contact Zoetis Customer Care

•  Incorrect test procedure (e.g., 

applying blood on the test strip 
before prompted by the meter)

•  There may be a problem with the 

test strip

• Meter error
• The test strip is affected by moisture

•  Wait for the drop and strip icons to appear before applying 

blood or control solution

•  Perform a control solution test using a new test strip. If the 

test results are within the range printed on the test strip vial, 
retest using blood and a new test strip

•  If the control solution result is out of range or the error 

reappears, contact Zoetis Customer Care

•  There may be a problem with the 

test strip

• Meter error
• Strip Sensing path broken
• Strip Type (OEM) error

 •  Perform a control solution test using a new test strip. If the 

test results are within the range printed on the test strip vial, 
retest using blood and a new test strip

•  If the control solution result is out of range or the error 

reappears, contact Zoetis Customer Care

Display message What it means

What to do

•  There may be a problem with the test 

strip connection with the meter

• Meter error

•  Perform a control solution test using a new test strip. If the 

test results are within the range printed on the test strip vial, 
retest using a fresh blood sample and a new test strip

•  If the control solution result is out of range or the error 

reappears, contact Zoetis Customer Care

No calibration code

Use the code strip attached to the test strip vial to set 
the correct lot code.

Code Strip reading failed

• Try to set the lot code again
•  If the code strip is illegible or broken, 

contact Zoetis Customer Care

Operating temperature too low

 Try again after warming up the device.

Error Codes (continued)

Содержание AlphaTrack3

Страница 1: ...1 2 User Guide 3 For In Vitro Diagnostic Use In Animals...

Страница 2: ...Blood Glucose Testing 14 9 0 Understanding Blood Glucose Results 18 Low and High Blood Glucose Results 18 10 0 Control Solution 20 Control Solution Information 20 Out of Range Control Solution Result...

Страница 3: ...meter is turned on You should check that the meter matches the example exactly every time the meter turns on Do not use the meter if the display check screen does not exactly match the example If the...

Страница 4: ...ure before using the AlphaTrak 3 meter Blood glucose monitoring should be done with the guidance of a veterinarian veterinary surgeon DO use the AlphaTrak 3 meter for Testing arterial blood Testing se...

Страница 5: ...e Mobile App helps pet owners to connect with their veterinarians to strengthen their relationship share their pet s glucose results and send reports It also allows veterinary professionals to easily...

Страница 6: ...the correct month then press m to save Day Press c to set the correct day then press m to save Time Format 5 Note The meter can display time in 12 hour e g 1 24P and 24 hour e g 13 24 time formats A i...

Страница 7: ...et to 65 mg dL 3 6 mmol L Note Hematocrit Hct warning is OFF by default Press c to turn the Hct warning ON or OFF then press m to save Press and hold m for 3 seconds until the meter turns off Bluetoot...

Страница 8: ...your hands before removing a strip from the vial Close the vial cap tightly immediately after removing a test strip Use the test strips beyond the expiration date printed on the side of the test stri...

Страница 9: ...Press c to set the correct code match the code on the strip vial Press m to save 5 Set Species Change the species on the meter after inserting the test strip Press and hold c and at the same time pre...

Страница 10: ...Discard used lancet and test strip as directed below Understanding Blood Glucose Results 8 View result The result will appear on the display when the test is complete 9 Add feeding information option...

Страница 11: ...to obtain control solution contact Zoetis Customer Care IMPORTANT Low or high blood glucose readings can indicate a potentially serious medical condition Consult a veterinarian regarding low and high...

Страница 12: ...tem check screen does not appear go to Troubleshooting How to Perform an AlphaTrak 3 control solution test continued 5 Set the correct species code See steps in Blood Glucose Testing section 6 Start c...

Страница 13: ...ote Two blinking arrows will indicate that results are being transmitted Press and hold m to exit meter memory Reminder Alarms 1 Press and hold m for more than 3 seconds to enter system settings 2 Pre...

Страница 14: ...ne after the other to avoid losing your settings If time and date settings are lost time and date will flash until reset Even if the time and date settings are lost you may still perform accurate bloo...

Страница 15: ...solution test using a new test strip If the test results are within the range printed on the test strip vial retest using blood and a new test strip If the control solution result is out of range or t...

Страница 16: ...or Codes continued Display message What it means What to do Dead flat battery Replace battery Reset date and time if necessary If meter still does not enter test mode contact Zoetis Customer Care Blue...

Страница 17: ...s not enter test mode after inserting a test strip Test strip is inserted upside down or bottom end in or not fully inserted into the meter Insert test strip print side up top end in No battery is ins...

Страница 18: ...an radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interferen...

Страница 19: ...ser d interf rences et 2 cet appareil doit accepter toute interf rence y compris les interf rences pouvant provoquer un fonctionnement ind sirable de l appareil Exposure to Radio Frequency Radiation C...

Страница 20: ...consequences Indicates important information about the use of the glucose meter Do not discard in household waste Needs to be sent to separate collection facilities for recovery and recycling DC curre...

Страница 21: ...etis Services LLC 333 Portage Street Kalamazoo MI 49007 USA All trademarks are the property of Zoetis Services LLC or a related company or a licensor unless otherwise noted u169 2021 Zoetis Services L...

Отзывы: