ECG MONITORING
740
SELECT
21-22-0335 Rev D
Page 21 of 40
HR PARAMETER SETUP
Table 11 on page 38 lists the Default HR Alarm Limits for Adult, Pediatric, and Neonatal.
on page 39 lists the Default ECG & HR Parameter Setting for Adult, Pediatric, and
Neonatal. HR Alarm Limits and Parameter settings will operate for the current monitor patient
mode.
HR ALARM LIMITS:
1) Touch HR Numeric field and the Setup HR menu will be displayed (refer to Figure 12);
Figure 12: HR Setup Menu
2) Set Alarms On to Yes to enable HR alarms, or No to disable HR alarms.
Warning:
If Alarms On is set
to “No” the monitor will not generate any visual or audible
indication of an alarm condition for HR.
3) Adjust the desired HR Upper or Lower Limit value.
The HR Upper and Lower Alarm Limits can be adjusted independently.
4) Touch OK to accept or Cancel to ignore the selection.
5) Touch the Home touch area to return to the Main screen.
HR AUTO (set) ALARM LIMITS
The Upper and Lower alarm limits for HR can be set automatically using the Auto feature,
which provides a structure to establish alarms limits based upon your patient’s current
measured value.
To enable Auto limits, open the HR
Setup Window and touch the “
AUTO
” key on the same
line with the parameter (refer to Table 1 for HR percentages).
Parameter
Alarm Limit Adjustment
(x% of the current measured value
Lower
Upper
HR
80%
125%
Table 1: HR Auto Alarm Limit Adjustment