SPECIFICATIONS
740
SELECT
21-22-0335 Rev D
Page 39 of 40
lists the ECG, HR & RR Parameter settings and Ranges.
Parameter
Defaults
Range
Adult
Pediatric
Neonatal
ECG Lead
II
II
II
I, II, III
(ECG Lead) Size
10 mm/mV
10 mm/mV
10 mm/mV
2.5, 5, 10, 15 mm/mV
HR Source
Auto
Auto
Auto
Auto, ECG, SpO
2
, NIBP
Pulse Tone Source
Off
Off
Off
ECG, SpO
2
, Off
ECG Filter
Monitor
(0.67
– 40 Hz)
Monitor
(0.67
– 40 Hz)
Monitor
(0.67
– 40 Hz)
Monitor (0.67
– 40 Hz),
Diagnostic (0.05
– 40 Hz)
Channel 1-3 Waveform
II
II
II
I, II, III
ECG Waveform Size
10 mm/mV
10 mm/mV
10 mm/mV
2.5, 5, 10, 15 mm/mv
Sweep Speed
25 mm/sec
25 mm/sec
25 mm/sec
6.25, 12.5, 25 mm/sec
Show Beat Detect Spikes
Yes
Yes
Yes
No, Yes
Notch Filter
60 Hz
60 Hz
60 Hz
60 Hz, 50 Hz, Off
Table 12: ECG, HR & RR Parameter Setting Ranges