4. RESPIRATION MONITORING
Note:
The following Warnings and Cautions are directed toward the Respiration monitoring
function. Additional Warnings and Cautions for the
740
SELECT
monitor are found in the
740
SELECT
User Manual, Zoe Medical PN 21-22-0316.
WARNINGS
Warning
:
Operating the
monitor with RR limit alarms disabled means that no low or high RR
alarm conditions will produce alarm notifications. Use this feature with extreme caution.
Patients must be closely observed if RR limit alarms are disabled.
Warning:
When using an ECG electrode to calculate respiration rate via the thorax
impedance method, movement artifacts may create inaccurate results. Respiration rates
derived from CO
2
parameter is not subject to such movement artifacts.
NOTES
Note:
If the Monitor is configured with a capnography module (ETCO
2
), the Monitor will not
allow the display of the Respiration Waveform or Numeric from the ECG source.
Refer to User Manual Addendum - ETCO
2
Parameters, Zoe Medical PN 21-22-0333 for
additional information regarding Respiration from ETCO
2
.
Note:
If the monitor is not configured for ETCO
2
, and Respiration waveform or Numeric is not
currently displayed, follow the instructions to enable the display the Respiration parameter in
the ENABLING DISPLAY OF RESPIRATION Section on page 27.
Note:
If the monitor is not configured for ETCO
2
, and the Respiration waveform is not
currently displayed, follow the instructions to display the Respiration waveform in the
RESPIRATION WAVEFORM section, page 28.
Note:
Configuring the monitor for Neonatal patient mode will disable the Respiration
parameter (Waveform and Numeric). Switching back to Adult or Pediatric patient modes will
require re-enabling the Respiration parameter.