ECG MONITORING
740
SELECT
21-22-0335 Rev D
Page 23 of 40
ANALYZE PACERS
Touch the
On
adjacent to the
Analyze Pacers
selection to Enable.
OR
Touch the
Off
adjacent to the
Analyze Pacers
selection to Disable.
When Analyze Pacers is set to On, the monitor will display white pace detect marks at the top
of the ECG waveform to indicate where pacer pulses were detected.
When Analyze Pacers is set to Off, the the monitor will display a message advising you that
the pacer detect function has been disabled.
Warning:
PACEMAKER PATIENTS
. When Analyze Pacers is disabled, the Heart Rate may
continue to count the pacemaker rate during occurrences of cardiac arrest. Keep pacemaker
patients under close surveillance and do not rely entirely upon Heart Rate alarms.
Figure 13: Analyze Pacers Enabled
PULSE TONE
The user-selectable
Pulse Tone
feature allows the clinician to select the desired to sound a
tone for each pulse detected.
User Selectable settings: Off (Default) & On
ECG LEAD
The user-selectable
ECG Lead
feature allows the clinician to select the desired ECG Lead
waveform on the display.
User Selectable settings: I, II (Default) & III
SIZE
The user-selectable
Size
feature allows the clinician to select the desired size of the ECG
waveform on the display.
User Selectable settings: 2.5, 5, 10 (Default), 15 & 20 mm/mV