(mWh/Wh) .
In the auto mode “charge-discharge-charge” , you can check testing result of each step by
quickly press “INC” or “DEC” button:
Press “On” button when you see testing interface to start testing and press it again to stop . In
the stop state press “SET” knob for two seconds or longer to open the setting interface (this
function disabled when the tester connects to a computer).
4.3 Setting interface
The cursor sets automatically on the testing mode and you can move it to the option you want
to change. One press on the “SET” button to move one step to right. Press “INC” or “DEC” to
change the parameter. The parameter will change automatically if parameters set exceed
limitation. Press the “ON” button to swith to next setting option.
1) Constant current discharge
First line: mode and discharge current
Second line: Cut-off voltage and maximum time
Discharge testing stops when the voltage reaches lower than cut-off voltage.
Testing time can be changed between 0-999 munites and to set it 0 means no limitation.
2) Constant power discharge
First line: mode and discharge power
Second line: Cut-off voltage and maximum time
Discharge testing stops when the voltage reaches lower than cut-off voltage.
Testing time can be changed between 0-999 munites and to set it 0 means no limitation.
3) Constant voltage charge