background image

   

                                                         

www.zketech.com 

 

         

EBC-A20 User Manual      

      Copyright (C) 2006-2017 ZKETECH 

3.2 Computer connection 

The  tester  can  be  connected  with  a  computer  through  a  designated  USB-TTL  cable  (Mini 
USB).   

3.3 Input Power   

The tester works on a power supply of DC 19-20V, 3.5A or above and interface specification of 
5.5/2.1 (inside positive and outside negative). 

Switch

Power

Comm

 

4. Display and setting 

4.1 Buttons 

ON   

—— Start

stop 

SET   

—— Set

switch 

 

INC    

—— Page up

increase 

 

DEC    —— Page down

decrease 

4.2  Testing interface 

  

 

The first line shows testing mode (CC/CP/CV), voltage and current.   

—

 

CC  —— constant current discharge 

—

 

CP  —— constant power discharge 

—

 

CV  —— constant voltage charge 

The second line shows working state (it shows OFF when the tester stops working and it turns 
DSC or CHG at working. It reads ATI when you choose auto mode charge-discharge-charge, 
and  1  refers  to  current  working  step),  working  time  (minutes)  and  capacity  (automatically 
switches between 0000mAh, 00.00Ah and 000.0Ah).   

A  short  press  on  the  “SET”  button  the  second  line  switch  to  show  energy  and  power 

Содержание EBC-A20

Страница 1: ...ZKETECH Released Date 2017 06 01 EBC A20 Battery Tester User Manual ...

Страница 2: ...tery capacity test DSC CP constant power discharge discharge power remains constantly the same recommended for simulation of constant power equipment Manual mode CHG CV constant voltage charge charge voltage remains constantly the same recommended for lithium and lead acid batteries test Auto mode The tester supports charge discharge charge auto circle tests recommended for capacity test Voltage a...

Страница 3: ...switch INC Page up increase DEC Page down decrease 4 2 Testing interface The first line shows testing mode CC CP CV voltage and current CC constant current discharge CP constant power discharge CV constant voltage charge The second line shows working state it shows OFF when the tester stops working and it turns DSC or CHG at working It reads ATI when you choose auto mode charge discharge charge an...

Страница 4: ... SET button to move one step to right Press INC or DEC to change the parameter The parameter will change automatically if parameters set exceed limitation Press the ON button to swith to next setting option 1 Constant current discharge First line mode and discharge current Second line Cut off voltage and maximum time Discharge testing stops when the voltage reaches lower than cut off voltage Testi...

Страница 5: ... charge we recommend you to set 5 10 minutes waiting time to cool batteries After completes settings press SET button for a long time to save data and return to the upper level interface Press SET button again to save data and return to the testing interface Press ON button to start tests 4 4 Setting examples 1 Setting steps Plug in power cable and turn on the tester Connect the tester with batter...

Страница 6: ...ge 3 For batteries with a voltage higher than 10V the maximum charging current should be lower than power current 0 5A 4 PC connection software EB Tester Software download it from www zketech com 5 PC connection follow EB Tester Software User Manual 6 Technical support tech zketech com 7 Business consultant sale zketech com We always focus on innovation and improvement and will keep upgrading the ...

Отзывы: