Oxygen Monitor SGM7.2 8 Operation
8.1.4 Status and error messages
During the measuring the functions of the measuring cell will be
monitored. In the disturbance-/ fault case error messages are
issued. The relay output will be activated simultaneously.
Own check of
the SGM7.2
Table 6: State-/Error messages
State
Display Remark
0 OK
1 LIMIT
2 RANGE
<<<
no alarm!
3 RANGE
>>>
no alarm!
4
FLOW <<<
< 5 l/h
5
FLOW >>>
>10 l/h
6
7
8
9
WARM UP
cell temperature too low (wait,
after approx. 15 minutes 750
°C has been reached)
10
CELL TEMP.<<<
set temp. -10° and > 30 min
11 ERROR
THERMOCOP.
thermo couple has broken
12
13
14
SELF TEST FAILED
Status 1;4...8:
warnings, only alarm relay active
Status 2,3:
warnings,
no alarm
by alarm relay
Status 9...14: failure, alarm relay active
and
analog output (voltage signal)