Problem

 

Cause

 

Solution

 

Eluate containing DNA   

is discolored. 

DNA contaminated with heme.  Minimize the amount of blood 

or blood-stained sample used 

(≤ 20 μl blood spot for forensic 

samples). 

DNA is sheared or degraded. 

Bubbles formed during mixing 
steps. 

 

To  prevent  bubble  formation 
during  mixing,  make  sure  the 
sample  volume  is  at  least  at 
the  recommended  volume 
listed in the resp. extraction kit 
handbook. 

Purified DNA repeatedly 
frozen and thawed. 

Aliquot purified DNA and store 
aliquots at 4°C (short-term) or 
-20°C 

(long-term). 

Avoid 

repeated 

freezing 

and 

thawing. 

DNA contaminated with 
DNase. 

Maintain a sterile environment 
while working (i.e., wear gloves 
and use DNase-free reagents). 

For any other technical issues/errors please contact your local distributor/agent. 

 

 
 
 
 
 
 
 
 
 
 

Содержание MagPurix 24 EVO

Страница 1: ...rp User Manual Automated Nucleic Acid Purification System 24 EVO Catalog No ZP01019 Manual No QW IH01 20 Version 3 0 Obelis s a Bd G n ral Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 73...

Страница 2: ...Safety Symbols Labels 5 1 3 2 About the Instrument 5 1 3 3 About Reagent Kits 6 1 3 4 About Samples 7 1 3 5 About Infectious Waste 7 2 Intended Use General Description 8 2 1 Intended Use 8 2 2 Externa...

Страница 3: ...Operation 19 5 1 Preparation 19 5 2 Start up the System 20 5 2 1 Basic Settings 20 5 2 2 User Management 20 5 2 3 Door Lock and Door Sensor Switch 20 5 3 Perform Extraction 21 5 4 System Fixing before...

Страница 4: ...www zinexts com MagPurix 24 EVO User Manual 3 35 9 2 Technical Support 34 10 Warranty Information 34 Revision History 35...

Страница 5: ...gPurix 24 EVO System is designed for the automated extraction and isolation of nucleic acids utilizing advanced magnetic bead separation technology 1 2 2 Applications The MagPurix 24 EVO System is int...

Страница 6: ...indicates that certain precautions must be taken when working with potentially infectious material Hot Surface This symbol labels potentially hot surfaces on the instrument Electricity This symbol is...

Страница 7: ...bjects on the MagPurix 24 EVO System Also refrain from bumping or knocking it as doing so may cause a failure or malfunction of the instrument If any liquid materials are left or have leaked inside th...

Страница 8: ...3 5 About Infectious Waste Caution When handling or disposing of infectious materials follow the laboratory guidelines or the law regarding infectious waste to perform proper incineration fusion steri...

Страница 9: ...ple types The MagPurix 24 EVO System is intended to be used only in combination with extraction kits indicated for use with MagPurix System for the applications described in the respective kit handboo...

Страница 10: ...www zinexts com MagPurix 24 EVO User Manual 9 35 2 3 Internal Features...

Страница 11: ...arts MagPurix 24 EVO System x 1 Power Cord x 1 Sample Rack for 2 ml Sample Tube x 2 Barcode Reader x 1 User Manual x 1 SD Card 32GB x 1 64GB x 1 Fuse x 3 1A 2A 4A Please make sure all components are f...

Страница 12: ...up and down suddenly Warming a cold room suddenly or moving the MagPurix 24 EVO System from a room with low temperature to a warm room may cause condensation inside the device resulting in abnormal op...

Страница 13: ...might lead to instrument damages For correct and safe use of the MagPurix 24 EVO System install the device in a location close to a power socket and with enough space for installation procedures and...

Страница 14: ...required for extraction are provided with the respective extraction reagent kits Contents of the kits will vary please refer to the resp extraction kit handbook for details 3 3 4 User Training User t...

Страница 15: ...wering up the MagPurix 24 EVO System the MagPurix 24 EVO System will start the initialization process Click LOGIN and User ID and Password will appear for input Set up the local time System will start...

Страница 16: ...nual 15 35 Camera time is suggested to be set for the first time use Follow the steps below to set the time Step 1 Remove the Camera as below Step 2 Turn on the instrument and click on Settings System...

Страница 17: ...35 Step 4 Find the buttons from the left side of the camera Step 5 Camera will start recording automatically after sensor is turned off Press OK button to pause Step 6 Long press M button to access to...

Страница 18: ...ser Manual 17 35 Step 8 Select Date Time and press OK Step 9 Use the button to set the time and press OK Step 10 Press M button to return to the main menu Step 11 On the panel click CAMERA to turn on...

Страница 19: ...access to users to certain options Once logged in you will be able to set up at least 10 other users with their own passwords 4 2 Main Screen The Main Screen of the MagPurix 24 EVO system on the righ...

Страница 20: ...he processed protocols and or some samples used with this instrument may contain infectious agents Caution Before start operating the instrument put on appropriate gloves a mask and safety goggles if...

Страница 21: ...are version installed is displayed at the bottom of the screen by clicking the Status Bar 5 2 2 User Management After logging in as administrator the Main Screen opens Click the Settings button and on...

Страница 22: ...MagPurix 24 EVO Software the Main Screen opens Click the Perform Run button and on the Perform Run Screen click the Extraction button Place reaction cartridges and all plastic consumables onto the wor...

Страница 23: ...er to separate reagent cartridge strips reagent chambers and tip holders if you do not run full batches of samples The position of Piercing Pin on the right and Filter Tip on the left are as follows T...

Страница 24: ...www zinexts com MagPurix 24 EVO User Manual 23 35 Note Zinexts does not guarantee with expired kit performance...

Страница 25: ...al To amend ID information DEL the wrong ID information and then scan the right ID Please take the Sample Tube Rack with the eluates and the samples out of the extraction module and discard all used e...

Страница 26: ...and 75 ethanol Clean the heating module with a soft paper tissue or soft cloth and use a cotton swab to remove any potential buffer residue from the hold of the heating module Run disinfection by tur...

Страница 27: ...tatus of the piston units Cleaning of piston UV light change annually Service Engineer 6 1 Cleaning of Rubbish Tray and M Motor Elevate M motor before the start of cleaning Clean the rubbish tray and...

Страница 28: ...24 EVO User Manual 27 35 6 2 Cleaning of Piston Enter Settings menu to select System and then click Piston Clean piston with 75 ethanol using soft cloth Change new piston if there is significant impur...

Страница 29: ...urns on when the power is on 1 The packing materials are not removed 2 Technical problem 1 Turn off the power and remove the packing materials 2 Contact your local representative or agent Protocol sto...

Страница 30: ...eparation of beads Too much sample material Reduce amount of sample material to the values recommended in the resp extraction kit handbook Carryover of magnetic beads into eluate Too much sample mater...

Страница 31: ...leakage from the tips 1 Perform preventive maintenance annually to ensure proper motor movements 2 Re load the consumables according to the instructions in this manual Low DNA yield Incomplete lysis D...

Страница 32: ...contain any particulate material that can clog the tip If needed centrifuge the sample prior to the extraction process No DNA recovered Magnetic beads stored or handled improperly Store cartridge cont...

Страница 33: ...event bubble formation during mixing make sure the sample volume is at least at the recommended volume listed in the resp extraction kit handbook Purified DNA repeatedly frozen and thawed Aliquot puri...

Страница 34: ...0 C assuming the room temperature at 25 C Protocol Input Touch Panel UV Light 30 minutes at 250 nm Built in Features 7 Touch Panel Instrument Dimensions 91 5 cm W x 66 5 cm D x 60 4 cm H Weight 100 kg...

Страница 35: ...at each product described herein will be free from defects in materials and workmanship for a period of twelve 12 months from the date of delivery Zinexts Life Science Corp agrees as its sole responsi...

Страница 36: ...ding on or obligate Zinexts Life Science Corp in any way In no event will Zinexts Life Science Corp be liable for any special incidental or consequential damages resulting from use or malfunction of t...

Отзывы: