31/40
2022.12.13 (Rev. 0)
[REMIFU-ZTN(EN)(NM)]
ZTone Instructions For Use
7. Messages
7.1 Over Temperature
The ZTone is an equipment that generates a magnetic field by applying a high current
to a transducer. As heat is generated from the transducer using the high current used to
create the magnetic field, cooling is performed by circulating the cooling oil inside the
transducer. If the transducer is used in a tilted position, the cooling oil will not reach all
sections of the transducer, and injury or stoppage of operation can occur. Because of
this, it is recommended to place the transducer in a horizontal position. If the transducer
overheats and the surface temperature reaches 40 degrees Celsius, the operation is
temporarily stopped and the message “Over Temperature” is displayed as shown in
[Figure 15]. When this message is displayed, put the transducer away from the person
for a while (Do not turn off the equipment) until the message disappears and the
equipment returns to a normal state.
Figure 15. Over temperature message
Caution
It is highly recommended to use the transducer in a horizontal
position.
Содержание ZTone non-medical
Страница 1: ...Instructions for Use ZTone non medical EN...