EMC Manufacturer's Declaration
17
Page 31
Medical electrical devices like Thermo
Pro
are subject to specific precautions relating to EMC (electromagnetic compatibility)
and must be installed and set up in accordance with the EMC information contained in the operating instructions or
accompanying documents.
Portable and mobile high-frequency communications equipment (e.g. mobile phones, cell phones) can affect medical
electrical equipment.
ThermoPro may only be operated with the original power cable indicated in the list delivered with the appliance.
Operating the device with a different power cable can result in increased emissions or decreased interference resistance of
the device!
Guidance and manufacturer's declaration – Electromagnetic emissions
The ThermoPro device is suitable for use in the specified electromagnetic environment. The purchaser or user of the
ThermoPro device should assure that it is used in an electromagnetic environment as described below:
Emissions test
Compliance
Electromagnetic Environment
RF emissions
CISPR 11
Group 2
The ThermoPro device must emit
electromagnetic energy in order to perform its
intended function. Nearby electronic equipment
may be affected.
RF emissions
CISPR 11
Class B
This ThermoPro device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
The device should not be used adjacent to or stacked with other equipment. When it is necessary to operate close to or
stacked with other equipment, the device should be checked to verify its correct operation in this arrangement.