Safety and Maintenance
14
Page 27
Thermo
Pro
is manufactured according to the safety regulations of DIN EN
60601-1.
This device has an expected service life of 10 years. As the device
manufacturer, Zimmer MedizinSysteme can support the safety and reliability of
this device through its service life if,
•
the user shall perform a periodic maintenance at least every four weeks to
ensure the safety of the device. Check the housing, the mains cable, the
applicator cable, and the applicator for any safety-related damage such as
gaps, cracks, and faulty isolation. Do NOT use the device and contact your
service representative if any damage is be found.
•
the device is operated at a proper power outlet with an grounding contact
and the electrical installation complies with DIN VDE 0100 part 710 or
similar,
•
the device is operated in accordance with the instructions for use,
•
extensions, readjustments or modifications are only carried out by persons
authorized by Zimmer MedizinSysteme,
•
the operator is satisfied regarding functional safety, proper condition and
mechanical integrity of the device and applicator prior to use of the device,
•
the device is operated only by properly qualified personnel,
•
the device is not operated in hazardous areas and/or a combustible
atmosphere,
•
the device is immediately disconnected from the mains in the event of
penetration of liquids.
The device does not contain any parts that can be serviced or repaired by the
operator.
No modification of this equipment allowed!
Regularly check the ventilation openings to ensure they are free of dirt. Clean if
necessary.
Note:
ThermoPro or its components do not require any periodic preventive
maintenance.