50
Electrotherapy
12.3 Description of the On/Off Time
button
12
On/Off Time active
Indicates that a On/Off Time is activated and shows the parameters of the
effective On/Off Time.
6.0 / 12.0 s
6.0 - rise time and retention time of the On/Off Time phase in seconds
12.0 - pause of the On/Off Time phase in seconds
After activation of the On/Off Time window, the On/Off Time parameters can
be set separately. For a detailed description see Section 10.5.
Note:
If the On/Off Time is activated during therapy, the therapy will be interrupted
and the intensity will be set to zero. The therapy time is stopped. The therapy
is restarted for the remaining time by adjusting the intensity.
On/Off Time
deactivated
When the On/Off Time is deactivated, the On/Off Time window is shown in
minimised form.
After activation of the On/Off Time window, the On/Off Time can be activated
and the On/Off Time parameters can be set separately.
For a detailed description see Section 10.5.
Note:
If there is no factory setting for a On/Off Time in a program, this window is not
shown.