Technical Information
10.2 Ultrasound
10
Page 28
Ultrasound heads
Frequency
0.8 MHz and 2.4 MHz
Ultrasound head, small 1 cm² , ERA = 0.67 cm² at 0.8 MHz,
0.65 cm² at 2.4 MHz
Maximum power
1.0 W at 0.8 MHz, 0.6 W at 2.4 MHz
Intensity levels
0.1 to 1 W/cm² eff. at intervals of 0.1 W / cm²
Ultrasound head, large 5 cm², ERA = 2.3
0
cm² at 0.8 MHz,
2.38 cm
2
at 2.4 MHz
Maximum power
6.9 W at 0.8 MHz, 7.1 W at 2.4 MHz
Intensity levels
0.1 to 3 W/cm² eff. at intervals of 0.1 W/cm²
Accuracy
± 20 %
Sound forms
1.
Continuous ultrasound
2.
Impulse noise, adjustable pulse frequencies:
20 Hz, 50 Hz, 100 Hz
Duty
ratio: 1: 1, 1: 2, 1: 3, 1: 5, 1: 10
Interchangeability
Ultrasound heads are calibrated ex works and can be easily replaced.
Appl
ied
P
art
Transducer
ultrasound head