SECTION
4
NexGen
®
Conventional Instruments – Tibial Technique
NexGen Conventional Instruments –
Tibial Technique
Position Tibial Pin Guide
•
Remove the meniscus.
•
Do not remove osteophytes from the tibia. Clean the bone
around the ACL and anterior tibia where the guide fits.
Remove any soft tissues that may prevent proper placement
of the Pin Guide.
•
Position the Pin Guide on the tibia
(Fig. 8)
.
•
The rotation cylinders do not make contact with bone surface.
Both cylinders will be above the tibia surface. All other
contact with the Pin Guide should ensure a proper fit.
•
The mechanical axis and proximal resection lines may be
used to asses the alignment. Additionally, the alignment
arch (00-5977-024-00) and rod (00-5785-080-00) may
be used to verify alignment. Additionally, the alignment
adapter (00-5970-024-00) and rod (00-5785-080-00)
may be used to verify alignment.
Drill and Pin Anterior Pin Holes
•
Drill and pin the medial and lateral anterior holes of the
Pin Guide
(Fig. 9)
.
Fig. 8
Fig. 9
1
B
Instruments
PSI Tibia
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
Alignment Adapter
00-5970-024-00
Alignment Rod
00-5785-080-00