Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation Surgical Technique
31
Finish the Patella
For the NexGen Primary Porous Patella
With Trabecular Metal Material
Center the appropriate Patellar Drill
Guide over the resected patella surface
with the handle on the medial side of
the patella and perpendicular to the
tendon. Press the drill guide firmly in
place so that the teeth fully engage and
the drill guide sits flat on the bone
surface (Fig. 7k). Drill the peg hole
making sure the drill stop collar contacts
the top of the drill guide (Fig. 7l).
NOTE:
The Primary Porous Patellar
Clamp may be used to fully seat
the drill guide on hard sclerotic
bone surfaces.
Fig. 7l
Fig. 7k
The depth gauge wing on the Patellar
Reamer Clamp can be used instead
of the stops to control the amount
of bone remaining, rather than the
amount of bone removed.
Insert the reamer assembly into the
Patellar Reamer Insetting Guide. Raise
the reamer slightly off the bone and
bring it up to full speed. Advance it
slowly until the prominent high points
are reamed off the bone. Continue
reaming with moderate pressure.
Remove the reamer clamp assembly.
Fig. 7m
For the NexGen
All-Polyethylene Patella
Center the appropriate Patellar Drill
Guide over the patella with the handle
on the medial side of the patella and
perpendicular to the tendon. Holding
the drill guide firmly in place, drill the
three peg holes using the Patellar/
Femoral Drill Bit (Fig. 7m).
Содержание MIS Multi-Reference
Страница 2: ......
Страница 45: ......
Страница 46: ......
Страница 47: ...Zimmer MIS Multi Reference 4 in 1 Femoral Instrumentation Surgical Technique 45 ...