System set-up
5
Page 7
Ensure that the device is placed on a stable surface.
No special training or instruction is required in order to use the device.
However, the device may only be used by medical professionals who have
previously familiarised themselves with the device and the instructions for use
and understand the intended use, application and warning notes outlined
therein.
Make sure that the power switch of the device is set to “0”.
Connect power cable
Connect the power cable to the provided port (9) on the device, and connect
the cable to the mains.
Note:
The device may only be connected to power outlets with an earthing contact.
Connect handpiece
Connect the handpiece to the handpiece connector (1).
Install gel pad
Only use the gel pads if they have been stored for at least 2 hours at room
temperature before therapy.
Rotate the gel pad fixation ring (18) anti-clockwise to remove it.
Select a gel pad and apply sufficient conductive gel to the back side of the
pad.
Place the back side of the gel pad into the handpiece, and put the gel pad
fixation ring (18) back on the handpiece
Fix the gel pad fixation ring (18) in place by rotating it clockwise.
Make sure to use sufficient conductive gel on the front side of the gel pad as
well as when performing a treatment.
Connect foot switch
The optional foot switch must be connected to the foot switch connector (4)
The foot switch has 3 buttons. The white button (left side) is used to apply the
set power. The middle button adjusts the energy downwards (blue). The
yellow button (right side) is used to adjust the applied energy upwards.
Switching device off
The device is switched off using the on/off switch (10).
In order to fully disconnect the device from the mains, the power cable must
be disconnected.
Caution
!
All cables must be protected from pinching or other mechanical damage.