Warnings
3
Page 5
Treatment instructions
regarding the location, duration and intensity of the
treatment require medical knowledge and should only be given by medical
professionals. These instructions must be followed.
The patient must not be left unattended during therapy.
Performing intracranial, transcardiac and cervical-occipital treatments is
prohibited. Failure to follow this instruction may lead to hazards to the patient.
Use in wet areas is not permitted. Non-compliance may lead to considerable
damage to the device and endanger both the patient and the user.
Never use the handpiece without an appropriate gel pad or with a damaged gel
pad. Failure to follow this instruction may lead to hazards to the patient.
Patient body parts which contain (metal) implants (for example, a bone nail,
pacemaker, ICD, IUD, etc.) must be excluded from the treatment.
Cables leading to the applicator should be guided in such a way that contact
with the patient or with conductive or energy-absorbing objects is avoided.
Ensure that the handpiece and gel pad are never directed towards the eyes
during operation.
Ensure that the handpiece and gel pad are not directed at metallic surfaces
(such as a couch, washbasin, device housing, etc.).
Ensure that the device is not opened.
Ensure that the power cable is immediately disconnected from the mains if
liquid or foreign bodies penetrate the housing.
Ensure that the device is inspected by an authorised service employee before it
is put back into operation.
To completely disconnect the device from the mains, pull the power cable out of
the socket.
Ensure that the ventilation slots are not covered. Do not cover the ventilation
slots with objects.