Application Information
3
Page 3
Before using the device on patients, users should familiarise themselves with
the operating instructions and individual treatment methods to be used as well
as the indications/contraindications and warnings/instructions for use.
Additional sources of information about the treatment should also be observed.
Caution
!
Before use, ensure that the device is operated via a properly grounded socket
(electrical installation in accordance with DIN VDE 0100 Part 710). The device
may only be operated with the supplied power cable. The power cable must be
protected against mechanical stress.
Caution
!
The operation of devices with strong electromagnetic fields (e.g. tomography,
X-ray, or diathermy equipment) may interfere with the operation of the device.
Please maintain a safe distance of several metres.
The enPulsPro is not suitable for use in explosive, flammable, or combustive
environments.
During use, the device must be located in a position that allows direct access to
the mains power supply so that this can be disconnected at any time.
To prevent the risk of electric shock, the plug must be disconnected from the
power supply before performing any maintenance or cleaning activities.
Inspect the device before use. If there is any damage, do not use the device.
Caution
!
Only use accessories from Zimmer MedizinSysteme GmbH.
Caution
!
The handpiece of the device is not designed for continuous operation. After a
max. 6,000 strokes, a treatment pause of 15 min is required.
Caution
!
If the enPulsPro is not mounted on a designated wagon, ensure that the
enPulsPro is placed on a stable surface.
Caution
!
To avoid heat accumulation in the handpiece, ensure that the vents on the top
and especially the bottom of the handpiece are not blocked (i.e. by the hand or
otherwise).