Manufacturer´s EMC declaration
18
Page 34
Guidelines and manufacturer declaration - electromagnetic interference resistance
The cl
Trac
device is intended for operation in the electromagnetic environment indicated below. The cl
Trac
customer or user
should ensure that it is used in such an environment.
Interference resistance
tests
IEC 60601 test level Compliance
level
Proximity fields from RF wireless
comunications equipment
IEC 60601-1-2:2014
Table 10
385 MHz, 27 V/m, Pulse 18 Hz, Horizontal, 5 sec.
450 MHz, 28 V/m, FM +5Hz, 1kHz sine, Horizontal, 5 sec.
710 MHz
745 MHz, 9 V/m, Pulse 217 Hz, Horizontal, 5 sec.
780 MHz
810 MHz
870 MHz, 28 V/m, Pulse 18 Hz, Horizontal, 5 sec.
930 MHz
1720 MHz
1845 MHz, 28 V/m, Pulse 217 Hz, Horizontal, 5 sec.
1970 MHz
2450 MHz, 28 V/m, Pulse 217 Hz, Horizontal, 5 sec.
5240 MHz
5500 MHz, 9 V/m, Pulse 217 Hz, Horizontal, 5 sec.
5785 MHz
385 MHz, 27 V/m, Pulse 18 Hz, Vertical, 5 sec.
450 MHz, 28 V/m, FM +5Hz, 1kHz sine, Vertical, 5 sec.
710 MHz
745 MHz, 9 V/m, Pulse 217 Hz, Vertical, 5 sec.
780 MHz
810 MHz
870 MHz, 28 V/m, Pulse 18 Hz, Vertical, 5 sec.
930 MHz
1720 MHz
1845 MHz, 28 V/m, Pulse 217 Hz, Vertical, 5 sec.
1970 MHz
2450 MHz, 28 V/m, Pulse 217 Hz, Vertical, 5 sec.
5240 MHz
5500 MHz, 9 V/m, Pulse 217 Hz, Vertical, 5 sec.
5785 MHz