Device Set Up
6
Page 7
Installing the traction
device
After the transport and before switching on the device, make sure that the device is
properly fixed to the traction table. Always follow the installation instructions from the
traction table mounting bracket. If not present, follow the instructions below:
Function of holes on device baseplate:
A.
Alignment hole with ø9.6mm
B.
Alignment set pin holes with ø6.5mm
C.
M8 Screw-holes for couch mount
D.
M8 Screw holes for stand mount or mounting bracket
Only a traction bank provided or recommended by Zimmer MedizinSysteme GmbH
must be used.
Install device on
therapy table /
mounting bracket
Hole A of the device fix on pin A of the installing plate (only on specific traction tables
available)
Turn the device to setting position as intended (there are 3 positions for setting)
Fix the cl
Trac
to the traction table plate by 4 M8 hand-screws.