SERVICE INFORMATION

Caution:

Risk of electric shock. Do not remove cover.

Return to an authorized Zimmer Orthopaedic Surgical
Products service center for servicing. Do not attempt to
disassemble the handpiece. It is a factory sealed unit with
no user serviceable parts inside. Refer servicing to
qualified service personnel.

DO NOT

attempt to adjust the power supply. It has been

factory calibrated to provide optimum cutting performance
and maximum safety. It has been designed to meet or
exceed specific medical electrical safety standards.

The 

Zimmer

Tissue Bank Dermatome should be returned

to an authorized Zimmer Orthopaedic Surgical Products
service center every 6 months for inspection, preventive
maintenance, and calibration.

FUSES

Input fuses are located in the fuse drawer in the power
entry module at the rear of the power supply.

120 V~1.25 A time lag/delay

TROUBLESHOOTING GUIDE (TABLE 1)

WARRANTY (U.S.A.)

Zimmer Orthopaedic Surgical Products warrants that the

Zimmer

Tissue Bank Dermatome, along with all parts and

accessories, have been tested and inspected, and have
left the factory in proper working condition, free from
visible defects.

Zimmer Orthopaedic Surgical Products warrants to the first
purchaser of the 

Zimmer

Tissue Bank Dermatome, Width

Plates, Screwdriver, and Power Supply that these products
and accessories will, under normal and reasonable use,
be free from defects in material and workmanship for one
(1) year after the date of shipment from the factory. The

Zimmer

Tissue Bank Dermatome Blades are warranted to

be free from defects in material and workmanship upon
delivery. During the warranty period, Zimmer Orthopaedic
Surgical Products shall repair (or at its sole option replace)
the defective product or part without cost to the purchaser.
Defective parts replaced under this warranty shall become
the property of Zimmer Orthopaedic Surgical Products.
This warranty does not cover damage caused by misuse,
abuse, accident, neglect, or any uses not prescribed in
this manual. If the unit becomes defective because of mis-
use or abnormal conditions of operation, repairs will be
billed at our current rate.

ALL OTHER WARRANTIES, EXPRESS, IMPLIED, OR
STATUTORY, INCLUDING, BUT NOT LIMITED TO,
IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE, ARE
LIMITED IN TIME TO THE PERIOD OF THE WARRANTY
GRANTED HEREBY. SOME STATES DO NOT ALLOW
LIMITATIONS ON THE DURATION OF AN IMPLIED
WARRANTY, SO THE ABOVE LIMITATIONS MAY NOT
APPLY TO YOU.

Neither Zimmer Orthopaedic Surgical Products nor the
Zimmer distributor who sells the 

Zimmer

Tissue Bank

Dermatome is responsible for indirect, incidental, or
consequential damages. Some states do not allow the
exclusion of incidental or consequential damages, so the
above limitations or exclusions may not apply to you.

Zimmer

Tissue Bank Dermatomes requiring service or

repair may be returned to Zimmer Orthopaedic Surgical
Products, 200 West Ohio Avenue, Dover, Ohio, U.S.A.
44622. A purchase order and return goods authorization
number must accompany all equipment for repair. The
customer will be responsible for all shipping charges. For
further information concerning warranties and repairs,
contact your Zimmer sales representative or the Zimmer
Customer Service Department at 800-348-2759 (inside the
U.S.A.) or 574-267-6131 (outside the U.S.A.).

WARRANTY (Outside U.S.A.)

Please contact your local Zimmer representative for
warranty information.

PROBLEM

ACTION

DERMATOME NOT

CHECK INPUT FUSES.

OPERATING.

CHECK POWER CORD

POWER LIGHT OFF.

CONNECTIONS.
RETURN FOR SERVICE.

DERMATOME NOT

CHECK BLADE FIT.

OPERATING.

CHECK HANDPIECE

POWER LIGHT ON.

CONNECTION TO
POWER SUPPLY.
RETURN FOR SERVICE.

DERMATOME

CHECK HANDPIECE

OPERATING TOO

CONNECTION TO

SLOWLY.

POWER SUPPLY.

ERRATIC SPEED

CHECK HANDPIECE

CHANGES.

CONNECTION TO
POWER SUPPLY.

ACCESSORIES 

ZIMMER

TISSUE BANK DERMATOME

See Fig. 16–23.

For further information contact your Zimmer sales 
representative or the Zimmer Customer Service
Department at 800-348-2759 (inside the U.S.A.) or 
574-267-6131 (outside the U.S.A.).

Содержание 8831-001

Страница 1: ...receiving the Zimmer Tissue Bank Dermatome examine the unit for external signs of damage Retain all packing material until the contents have been verified and an initial operational check has been pe...

Страница 2: ...CAUTIONS The on off lever must be in the SAFE position before changing blades when the instrument is not in use or when connecting power to the instrument Accidental activation of the instrument durin...

Страница 3: ...ply Safely dispose of all used blades Thoroughly scrub the dermatome and width plate s with a soft brush and mild detergent All traces of blood coagulated material stains etc should be removed Rinse a...

Страница 4: ...amage caused by misuse abuse accident neglect or any uses not prescribed in this manual If the unit becomes defective because of mis use or abnormal conditions of operation repairs will be billed at o...

Страница 5: ...5 1 2 A Autoclave Case B Power Supply C Screwdriver D Width Plates E Handpiece A E D C B...

Страница 6: ...cle D Power Entry Module E Back SYMBOL DEFINITIONS A Type of Equipment B Direct Current C Year of Manufacture D OFF Power Disconnection from the Mains E On Power Connection to the Mains F Alternating...

Страница 7: ...7 Handpiece A On Off Safety Lock B On Off Lever C Handpiece D Connector Plug E Thickness Control Lever Blade Replacement A Screwdriver B Width Plate C Blade D Carriage 6 7 10 9 8 A A B C D B C D E...

Страница 8: ...8 11 15 12 METHOD I A Handpiece B Pocket METHOD II A Handpiece B Forceps 13 14 A B A B...

Страница 9: ...DESCRIPTION Zimmer Tissue Bank Dermatome Complete includes 8831 01 8821 06 8831 03 REF 8831 06 DESCRIPTION Power Supply REF 8831 03 DESCRIPTION Autoclave Case REF 8831 01 DESCRIPTION Handpiece includ...

Страница 10: ...831 000 10 DESCRIPTION Zimmer Tissue Bank Dermatome Blades Sterile 10 per box REF 8831 04 8831 05 DESCRIPTION Width Plates REF 8803 001 10 DESCRIPTION Width Plate Screws 10 per pack REF 8803 DESCRIPTI...

Страница 11: ...11...

Страница 12: ...Zimmer Orthopaedic Surgical Products 200 West Ohio Avenue P O Box 10 Dover Ohio U S A 44622 Revised 05 05 2005 Zimmer Inc Printed in U S A 18 10620 Printed on Recycled Paper...

Отзывы: