SERVICE INFORMATION
Caution:
Risk of electric shock. Do not remove cover.
Return to an authorized Zimmer Orthopaedic Surgical
Products service center for servicing. Do not attempt to
disassemble the handpiece. It is a factory sealed unit with
no user serviceable parts inside. Refer servicing to
qualified service personnel.
DO NOT
attempt to adjust the power supply. It has been
factory calibrated to provide optimum cutting performance
and maximum safety. It has been designed to meet or
exceed specific medical electrical safety standards.
The
Zimmer
Tissue Bank Dermatome should be returned
to an authorized Zimmer Orthopaedic Surgical Products
service center every 6 months for inspection, preventive
maintenance, and calibration.
FUSES
Input fuses are located in the fuse drawer in the power
entry module at the rear of the power supply.
120 V~1.25 A time lag/delay
TROUBLESHOOTING GUIDE (TABLE 1)
WARRANTY (U.S.A.)
Zimmer Orthopaedic Surgical Products warrants that the
Zimmer
Tissue Bank Dermatome, along with all parts and
accessories, have been tested and inspected, and have
left the factory in proper working condition, free from
visible defects.
Zimmer Orthopaedic Surgical Products warrants to the first
purchaser of the
Zimmer
Tissue Bank Dermatome, Width
Plates, Screwdriver, and Power Supply that these products
and accessories will, under normal and reasonable use,
be free from defects in material and workmanship for one
(1) year after the date of shipment from the factory. The
Zimmer
Tissue Bank Dermatome Blades are warranted to
be free from defects in material and workmanship upon
delivery. During the warranty period, Zimmer Orthopaedic
Surgical Products shall repair (or at its sole option replace)
the defective product or part without cost to the purchaser.
Defective parts replaced under this warranty shall become
the property of Zimmer Orthopaedic Surgical Products.
This warranty does not cover damage caused by misuse,
abuse, accident, neglect, or any uses not prescribed in
this manual. If the unit becomes defective because of mis-
use or abnormal conditions of operation, repairs will be
billed at our current rate.
ALL OTHER WARRANTIES, EXPRESS, IMPLIED, OR
STATUTORY, INCLUDING, BUT NOT LIMITED TO,
IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE, ARE
LIMITED IN TIME TO THE PERIOD OF THE WARRANTY
GRANTED HEREBY. SOME STATES DO NOT ALLOW
LIMITATIONS ON THE DURATION OF AN IMPLIED
WARRANTY, SO THE ABOVE LIMITATIONS MAY NOT
APPLY TO YOU.
Neither Zimmer Orthopaedic Surgical Products nor the
Zimmer distributor who sells the
Zimmer
Tissue Bank
Dermatome is responsible for indirect, incidental, or
consequential damages. Some states do not allow the
exclusion of incidental or consequential damages, so the
above limitations or exclusions may not apply to you.
Zimmer
Tissue Bank Dermatomes requiring service or
repair may be returned to Zimmer Orthopaedic Surgical
Products, 200 West Ohio Avenue, Dover, Ohio, U.S.A.
44622. A purchase order and return goods authorization
number must accompany all equipment for repair. The
customer will be responsible for all shipping charges. For
further information concerning warranties and repairs,
contact your Zimmer sales representative or the Zimmer
Customer Service Department at 800-348-2759 (inside the
U.S.A.) or 574-267-6131 (outside the U.S.A.).
WARRANTY (Outside U.S.A.)
Please contact your local Zimmer representative for
warranty information.
PROBLEM
ACTION
DERMATOME NOT
CHECK INPUT FUSES.
OPERATING.
CHECK POWER CORD
POWER LIGHT OFF.
CONNECTIONS.
RETURN FOR SERVICE.
DERMATOME NOT
CHECK BLADE FIT.
OPERATING.
CHECK HANDPIECE
POWER LIGHT ON.
CONNECTION TO
POWER SUPPLY.
RETURN FOR SERVICE.
DERMATOME
CHECK HANDPIECE
OPERATING TOO
CONNECTION TO
SLOWLY.
POWER SUPPLY.
ERRATIC SPEED
CHECK HANDPIECE
CHANGES.
CONNECTION TO
POWER SUPPLY.
ACCESSORIES
ZIMMER
TISSUE BANK DERMATOME
See Fig. 16–23.
For further information contact your Zimmer sales
representative or the Zimmer Customer Service
Department at 800-348-2759 (inside the U.S.A.) or
574-267-6131 (outside the U.S.A.).