AccuMix Mixing System: AccuFill Mixing Technique
1
Transfer the tray to the
sterile field (back table).
Remove the mixing syringe
and set upright in the
tray groove; lift funnel to
extend syringe.
7
Remove mixing syringe
from tray. Remove plunger
sleeve from plunger stem.
DO NOT DISCARD SLEEVE!
8
Thoroughly mix powder
and saline for 60 strokes
(~60 seconds). Twist and
rotate while plunging
until mix takes “paste”
consistency.
9
Remove blue plug.
Reattach sleeve to stem,
with stem fully extended.
10
Holding syringe with cap
upright, expel excess air
from syringe. Connect the
first 1 cc syringe. Inject
AccuFill BSM into syringe.
Repeat for remaining
syringes.
11
Transfer filled syringes
to operative field.
2
Remove vial of AccuFill
powder from jar. Empty
powder into funnel;
tap until powder enters
syringe.
3
Remove funnel; fully
tighten cap and plug.
Remove blue plug and
set in sterile tray.
DO
NOT DISCARD PLUG!
SET UP
MIX
HYDRATE
TRANSFER
4
Withdraw saline from vial using the saline
syringe and adaptor.
P/N 201.150 (5 cc AccuFill BSM)
• 3.0 cc Saline
• Alternative: 3.4 cc Blood
P/N 201.130 (3 cc AccuFill BSM)
• 2.0 cc Saline
• Alternative: 2.3 cc Blood
5
Connect Saline syringe to white cap; tighten.
Inject Saline into powder; pull up on syringe
plunger to pull excess air into saline syringe.
Inject again, to ensure ALL SALINE FLOWS
INTO POWDER, then repeat to release
pressure.
6
Remove saline syringe; set it in the sterile
tray. Attach blue plug to cap.
NOTE: Do not empty entire contents of saline vial into
AccuFill BSM powder. Measure and use only the exact
volume noted above.
All content herein is protected by copyright, trademarks and
other intellectual property rights, as applicable, owned by or
licensed to Zimmer Biomet or its affiliates unless otherwise
indicated, and must not be redistributed, duplicated or
disclosed, in whole or in part, without the express written
consent of Zimmer Biomet. This material is intended for
health care professionals and the Zimmer Biomet sales
force. Distribution to any other recipient is prohibited. Not
for distribution in France. For product information, including
indications, contraindications, warnings, precautions,
potential adverse effects and patient counseling information,
see the package insert and www.zimmerbiomet.com. Check
for country product clearances and reference product specific
instructions for use.
AccuFill
®
BSM is Manufactured by ETEX Corporation,
38 Sidney Street, Cambridge, MA 02139 USA • etexcorp.com
AccuMix
®
is Manufactured by: Zimmer Knee Creations,
841 Springdale Drive, Exton, PA 19341 USA
©2017 Zimmer Biomet
1102.1-EMEA-en-REV0517
Authorized Representative
Zimmer GmbH
Sulzerallee 8
8404 Winterthur
Switzerland
Legal Manufacturer
ETEX Corporation
38 Sidney Street,
Cambridge, MA 02139
USA
Legal Manufacturer
Zimmer Knee Creations
841 Springdale Drive
Exton, PA 19341
USA
www.zimmerbiomet.com