TR250
12410-0701-03
Page
16 / 18
www.ziehl.de
Sensor connection :
Pt 100, Pt 1000 acc. to EN 60751
Range °C
Short-circuit
Ohm
Break
Ohm
Resistance of
line
Ohm
Sensor
min
max
<
>
max
Pt 100
-199
860
15
400
500
Pt 1000
-199
860
150
4000
4100
KTY 83
-55
175
150
4000
4100
KTY 84
-40
250
150
4000
4100
PTC
20
20000 *
* when no value 3800 ... 20000 ohm has
been measured before
Accuracy
±0,5 % of value ±1 K (KTY ±5 K)
Sensor-current
≤ 1 mA
Measuring time t
M
<2,5 s (depending on number and type of
connected sensors)
Housing:
type V 4
Mounting height
55 mm
Width
4 TE
Dimensions (W x H x D)
70 x 90 x 58 mm
Line connection solid wire
each 1 x 1,5 mm
2
Stranded wire with insulated ferrules
each 1 x 1,0 mm
2
Torque
0.5 Nm (3,6 lb.in)
Protection class housing
IP 30
Protection class terminals
IP 20
Fitting position
any
Mounting
Snap mounting on 35 mm standard rail
EN 60 715 or M4 screws
Weight
app. 200g
Subject to technical modifications