FR(MU)1000
12460-0751-02
Seite
8 / 8
www.ziehl.de
Cable length for proximity switch
PNP, NPN, 2-wire
Resistance of line
≤ 10 Ω / line
Capacity of line
≤ 22 nF 0…800Hz; ≤ 10 nF 800...1600 Hz
e.g. max. length of cable
< 150 m with cable LIFYY11Y 3*0,34 mm
at 0…800Hz
Measuring error
± 0,05 % of measured value ± 1 Digit
Temperature factor
< 0,002 %/K
Measuring time
1 Period *
Svm
(number mean values)
>= 3 Periods; after placing the measuring signal
Auxiliary 18 V 20 mA
16 ... 21 V max. 20 mA
Enable E1-E2
18 V / 3 mA Switching threshold approx. 9 V
Analogue output: (FRMU only)
electrically insulated to input f - in (U1-U2)
Voltage output 0...10 V
max. 10 mA error <0,1 % of full scale
Temperature factor
< 0,01 %/K
Current output 0/4...20 mA
max. 500 Ω error <0,15 % of full scale
Temperature factor
< 0,015 %/K
Error from burden
(250 Ω - resistance)/250 Ω * 0,15 % of final value
Nominal rise time
<20ms + messuring time
Resolution analog output
>= 11,6 bit
Subject to technical changes
9