11.1 Description of Menu Items :

11.1.1 Setting the GoNoGo values :

(at all ranges except V     )

Operating Order :

Basic

noGo

nnnn

Display
Setting :

nnnn = noGo limit values in M

(for Insulation ranges)

W

nnnn = noGo limit values in

(for Ohm and Continuity)

W

1.Now scroll through different values using the                key until

you come across required value.

2.To save this value press                 key. After saving, this will take

you to basic display.

3.To go back without saving press "ESC" key.This will take you to

main menu "noGo"

4.To go back to basic display from main menu "noGo" use "ESC" key.

11.1.2 Setting the Timer values

(Only at Insulation Resistance measurement ranges)

Operating Order :
Basic

noGo

ti                        mm:ss

display

Setting :

mm= minutes, ss = seconds

1. Now scroll through different values using the               key until

you come across required value.

2. To save this value press              key. after saving, this will take

you to basic display.

3. To go back without saving press

key.This will take you to

"ESC"

main menu "ti"

4. To go back to basic display from main menu "ti" use

key.

"ESC"

11.1.3 To see the stored values

(at all ranges except V      )

Operating Order :
Basic

noGo

ti              reCL               nnnn

display

1. You can now see the first stored value.The scale pointer is used

to indicate the number (out of 10) of the displayed stored value.

2. Now you can view the successive stored value using the

key.The scale pointer also increments to indicate the number
(out of 10) of the displayed stored value.

3. To go back press

key.This will take you to

"ESC"

main menu "reCL"

4 To go back to basic display from main menu "reCL" use

key.

"ESC"

Table for default values :

NoGo

ti

reCI

50V

100V

250V

500V

1000V

Ohm

Cont

V

1 M

W

1 M

W

5 M

W

5 M

W

500 M

W

00

W

5

W

---

00:00 (Manual)

---

---

---

0.00

0.00

0.00

0.00

0.00

0.00

0.00

---

www.ziegler-instruments.com

8

| 10

12. Specifications

Meas.

Range

Resolu-

Accuracy

Overload

function

tion

(..% of rdg

value and

+ .. digit)

duration

±

1) For insulation resistance range, 2) For low Ohms/continuity Ranges
*

Terminal voltage on open

* Open circuit voltage

circuit (DC)

5V  ± 1V d.c.

- 0% to + 30% of rated voltage * Lead resistance

*

Short circuit current < 2 mA

compensation : 0 - 9.99

W

*

Test current on load 1 mA at
minimum pass values of
insulation as specified
in VDE 0413 Part 1.

Reference Conditions :

0

Ambient Temp.

+23C ± 2 K

Relative Humidity

45%...55%

Battery Voltage

8V ± 0.1V

Voltage Measurement

AC (Sine), 50/60 Hz

Display

LCD display field (65 mm x 30 mm) with analog indication and Digital
display and with display of unit of measured quantity and functions

Analog

Display

LCD scale with bar graph pointer

Scale Length

47 mm

Scaling

0...30 with 30 graduations

Over range Indication

By triangle (13)

Digital

Display/Char. Height

7 segment digits/12mm

Number of Digits

3 digit for M

and

W

W

4 digit for Stop watch

Over range Indication

OL

Insulation
Resistance

1)

M

W

U = 50 V,

N

100 V

0.01 M

to

W

0.99 M

W

³

W

1.0 M

to

9.9 M

W

³

W

10 M

to

99 M

W

10 K

W

(

)

0.01 M

W

100 K

to

W

(

)

0.1M

W

1 M

W

± 3% ± 2

± 5% ± 2

± 30%

1200V rms,

10 sec

Insulation
Resistance

1)

M

W

U = 250 V,

N

500 V,
1000 V

0.01 M

to

W

9.99 M

W

³

W

10.0 M

to

99.9 M

W

³

W

100 M

to

999 M

W

10 K

W

(

)

0.01 M

W

100 K

to

W

(

)

0.1M

W

1 M

W

± 3% ± 2

± 5% ± 2

± 30%

1200V rms,

10 sec

2)

Low Ohms

W

0

9.99

-

W

³ W-

W

10

99.9

0.01

at

W

210 mA

0.1 W

at

21 mA

± 3% ± 2

± 5% ± 2

1200V rms,

10 sec

2)

Continuity

0

9.99

-

W

³ W-

W

10

99.9

0.01

at

W

210 mA

0.1 W

at

21 mA

± 3% ± 2

± 5% ± 2

1200V rms,

10 sec

V AC/DC

1 V

± 2% ± 3

1200V rms,

10 sec

³

-

25V

450V

³

-

450V

600V

1 V

± 3%

Содержание RI 10

Страница 1: ...15030858 www ziegler instruments com 1 10 Ziegler Instruments Order No E1 R0 921022 2 2014 EN RI 10 User Manual RI 10 Analog Digital Insulation Tester User Manual Analogue Meters With Moving Iron Movement ...

Страница 2: ...On Off push button selected functions 3 Push button 11 Display for selected function 4 Push button 12 Display for the unit of 5 TEST Push button measured quantity 6 Function selector switch 13 Over range Indication 7 Terminal sockets 14 Pointer for Analog indication 8 Symbol for 15 Scale for Analog Indication CONTINUOUSLY ON 16 Activated stop watch Indicator 9 Digital display with indication 17 Ze...

Страница 3: ...CE EU Confirmity mark Repair Parts Replacement and Calibration After opening the instrument live parts may be exposed Therefore the instrument must be disconnected from the measuring circuit prior to opening its case for repair replacement of parts or calibration If repair or calibration cannot be avoided unless the instrument is open and live the work must be performed by a qualified person who u...

Страница 4: ...t probes after insulation resistance measurement has been activated at the instrument A current with a value of 2 mA limited by the instrument may flow over your body and although this is not life endangering the electric shock is distinctly perceptible If you are taking a measurement at a capacitive device under test for example a cable it may be charged with as much as 1000V depending upon the s...

Страница 5: ...each range of insulation resistance measurement as well as for Ohms Continuity ranges Refer to the 11 Menu structure section for how to set the value for the noGo value 6 2 Storing the measured resistance value l To store the value of resistance press STORE key when the resistance measurement is in progress This will store the resistance measured at the instant the key is pressed l This way maximu...

Страница 6: ...d resistance value will be displayed on the display The instrument generates a continuous sound signal if the measured resistance value is less than the pre setted GoNoGO value Default value is 0W Refer to the 11 Menu structure section for how to set the value for the limit value 7 3 Zero adjustment When measuring very low resistance values you can eliminate resistance of leads and contact resista...

Страница 7: ... and Continuity Tests are inhibited When external voltage 25 V is present only the Insulation Tests are available Note The default live circuit warning does not operate i when in STOPWATCH mode ii when in MENU structure RI 10 M W 50V 100V 250V 500V 1KV V www ziegler instruments com 7 10 10 Stop Watch This function allows you to measure elapsed time up to one hour To use stop watch 1 Set the functi...

Страница 8: ...Table for default values NoGo ti reCI 50V 100V 250V 500V 1000V Ohm Cont V 1 MW 1 MW 5 MW 5 MW 500 MW 00 W 5 W 00 00 Manual 0 00 0 00 0 00 0 00 0 00 0 00 0 00 www ziegler instruments com 8 10 12 Specifications Meas Range Resolu Accuracy Overload function tion of rdg value and digit duration 1 For insulation resistance range 2 For low Ohms continuity Ranges Terminal voltage on open Open circuit volt...

Страница 9: ...ent which may remain unused for extended period of time Auto turn OFF Instrument turns off automatically if any of the keys or the function selector switch have not been activated for about 10 minutes in insulation resistance range and 5 minutes in other ranges Cable Set 2 Cu cross sectional area 1 mm End metallic parts should pass terminal 4 D 11 Technical data as per IEC 61010 EN 61010 2 031 CAT...

Страница 10: ...he overload before placing the instrument back in to operation Fuse Testing Set the function selector switch to any insulation measurement range Open circuit the test leads Press the Test button until OL or FUSE appears on display If display OL Fuse is OK If display FUSE Fuse is blown and needs replacement Note Once FUSE is displayed all the functions are inhibited Immediately switch the Instrumen...