5.2.2 Continuous measurement for a pre-selectable time :

In normal course, the insulation test terminates when the test key
is released and the measured insulation resistance value remains
on display

for few seconds

after the test key

is released.

With

the

Pre-selectable

measurement

time

feature,

the

insulation test continues for the pre-determined time.
Pre-selectable time : 10 sec-5 min
Refer to the “11. Menu structure” section for how to set the value for pre-
selectable time. Once the time is set it is applicable to all the 5 ranges of
Insulation measurement.

n

Select the required value of the time for which the test is
to be carried out.

n

Briefly Press the TEST key.

The insulation resistance

test carries on for the time interval selected.

n

The test terminates automatically after the selected time
has elapsed.

n

The display then flashes the value of measurement at
the end until the           Key is pressed.

5.3  Conclusion of measurement and discharging

After the measurement has been completed, any residual voltage is
displayed which may be caused by capacitive circuits. Contact with the
device under test must be maintained.The reduction in voltage can be
directly observed at the LCD.

Do not disconnect the device under test until voltage has dropped
below 25 volts !!

Note : Automatic discharge !

The capacitive circuits are automatically discharged when
TEST button is released following an insulation resistance test.
This auto discharge facility eliminates the need for the user to
discharge the circuit under test manually.

Additional features available in Insulation measurement ranges.

1.

Pre-selectable limit checks (Go/Nogo option)

2.

Storing the measured resistance value

3.

Minimum and Maximum value “MIN/MAX” storage

www.ziegler-instruments.com

5

|

10

6 Additional Features

6.1

Pre-selectable limit checks (Go/No-Go option)

An acoustic signal can be generated when the measured value of
resistance falls below an adjustable noGo value.
A separate noGo value can be set for each range of insulation
resistance measurement as well as for Ohms, Continuity ranges.
Refer to the “11. Menu structure” section for how to set the value
for the noGo value.

6.2

Storing the measured resistance value

l

To store the value of resistance press “STORE” key when the
resistance measurement is in progress. This will store the
resistance measured at the instant the key is pressed.

l

This way maximum 10 readings per range can be stored.

th

l

After 10 values are stored for the selected range the 11

value

when stored will over-write the first value i.e. at any given point
of time only ten readings per range can be stored.

l

The readings are stored in a volatile memory.Therefore when the
instrument is switched off, the readings are lost.

Recalling the stored values

The stored values can be recalled and viewed in the menu
structure.
Refer to the “11.Menu Structure” section for how to view the
stored value.

6.3

Minimum value and Maximum value “MIN/MAX” storage
facility (for all measuring ranges)

With the MIN/MAX function, you can hold the minimum and the
maximum measured value which was applied to the input of the
instrument after activating MIN/MAX.The most important application is
determination of minimum and the maximum value for long term
monitoring of measured quantities.
The function MIN/MAX can be activated in all measuring ranges.
MIN/MAX does not influence analog indication, you can still read the
actual measured value.

Apply the measured quantity to instrument prior to activating MIN/MAX
function.

The MIN/MAX function is switched off when.
1.

Key is pressed for approximately for 1 Sec. OR

2. Then function selector switch is operated OR
3. When the instrument is turned OFF and ON again

Function

MIN/MAX

Measured Value

MIN and MAX

Meter reaction

Display

Measured value DigitalMIN/MAX

Sound

signal

1. Activate

and store

2. Store

and display

3. Return

to 1.

Reset

short

short

short

short

long

Stored

Stored continued
in background
new MIN and MAX
values are displayed
Same as 1.

Stored values are

not cleared

Cleared

Actual measured value

Stored MIN value

Stored MAX value

Same as 1.

Cleared

MIN/MAX

flash

MIN flash

MAX flash

Same as 1.

Cleared

1x

1x

1x

1x

2x

Содержание RI 10

Страница 1: ...15030858 www ziegler instruments com 1 10 Ziegler Instruments Order No E1 R0 921022 2 2014 EN RI 10 User Manual RI 10 Analog Digital Insulation Tester User Manual Analogue Meters With Moving Iron Movement ...

Страница 2: ...On Off push button selected functions 3 Push button 11 Display for selected function 4 Push button 12 Display for the unit of 5 TEST Push button measured quantity 6 Function selector switch 13 Over range Indication 7 Terminal sockets 14 Pointer for Analog indication 8 Symbol for 15 Scale for Analog Indication CONTINUOUSLY ON 16 Activated stop watch Indicator 9 Digital display with indication 17 Ze...

Страница 3: ...CE EU Confirmity mark Repair Parts Replacement and Calibration After opening the instrument live parts may be exposed Therefore the instrument must be disconnected from the measuring circuit prior to opening its case for repair replacement of parts or calibration If repair or calibration cannot be avoided unless the instrument is open and live the work must be performed by a qualified person who u...

Страница 4: ...t probes after insulation resistance measurement has been activated at the instrument A current with a value of 2 mA limited by the instrument may flow over your body and although this is not life endangering the electric shock is distinctly perceptible If you are taking a measurement at a capacitive device under test for example a cable it may be charged with as much as 1000V depending upon the s...

Страница 5: ...each range of insulation resistance measurement as well as for Ohms Continuity ranges Refer to the 11 Menu structure section for how to set the value for the noGo value 6 2 Storing the measured resistance value l To store the value of resistance press STORE key when the resistance measurement is in progress This will store the resistance measured at the instant the key is pressed l This way maximu...

Страница 6: ...d resistance value will be displayed on the display The instrument generates a continuous sound signal if the measured resistance value is less than the pre setted GoNoGO value Default value is 0W Refer to the 11 Menu structure section for how to set the value for the limit value 7 3 Zero adjustment When measuring very low resistance values you can eliminate resistance of leads and contact resista...

Страница 7: ... and Continuity Tests are inhibited When external voltage 25 V is present only the Insulation Tests are available Note The default live circuit warning does not operate i when in STOPWATCH mode ii when in MENU structure RI 10 M W 50V 100V 250V 500V 1KV V www ziegler instruments com 7 10 10 Stop Watch This function allows you to measure elapsed time up to one hour To use stop watch 1 Set the functi...

Страница 8: ...Table for default values NoGo ti reCI 50V 100V 250V 500V 1000V Ohm Cont V 1 MW 1 MW 5 MW 5 MW 500 MW 00 W 5 W 00 00 Manual 0 00 0 00 0 00 0 00 0 00 0 00 0 00 www ziegler instruments com 8 10 12 Specifications Meas Range Resolu Accuracy Overload function tion of rdg value and digit duration 1 For insulation resistance range 2 For low Ohms continuity Ranges Terminal voltage on open Open circuit volt...

Страница 9: ...ent which may remain unused for extended period of time Auto turn OFF Instrument turns off automatically if any of the keys or the function selector switch have not been activated for about 10 minutes in insulation resistance range and 5 minutes in other ranges Cable Set 2 Cu cross sectional area 1 mm End metallic parts should pass terminal 4 D 11 Technical data as per IEC 61010 EN 61010 2 031 CAT...

Страница 10: ...he overload before placing the instrument back in to operation Fuse Testing Set the function selector switch to any insulation measurement range Open circuit the test leads Press the Test button until OL or FUSE appears on display If display OL Fuse is OK If display FUSE Fuse is blown and needs replacement Note Once FUSE is displayed all the functions are inhibited Immediately switch the Instrumen...

Отзывы: