background image

Frequency, Resistance, Diode, Continuity , 

Capacitance Test 

250V DC/AC 

Temperature (

/

℉)

 

250V DC/ AC 

 

 

 

Meter Description 

1.  Current clamp 
2.  Non-contact AC voltage indicator light 
3.  Clamp trigger 
4.  Rotary Function swith 
5.  Data Hold/ Back Light button 
6.  Relative button 
7.  LCD display 
8.  MODE select button 
9.  Range button 
10. Hz% Hlod button 
11. COM input jack 
12. V 

Ω

CAP 

TEMPHz

 jack 

13. Battery Cover 

 

 
 

1. 

AC  DC 

     AC (alternating current)  

and DC (direct currrent) 

2. 

                  Minus sign 

3. 

8.8.8.8

 

  4000 count (0 to 3999)  

measurement reading  

4. 

AUTO 

       AutoRange mode 

5. 

REL 

          Relative mode 

6. 

                   Diode test mode  

7. 

   

•)))

          Audible Continuity 

8.

 

HOLD 

       Data Hold mode 

 

9.

 

,

,

 

,m,V,A,K,M,

Ω

,

 Units of measure list 

10.

 

Hz %              

Frequency/duty cycle test mode

 

 
 

Specifications

 

Function 

Range & Resolution 

Accuracy (% of reading) 

AC Current 
(50/60Hz) 

40.00 AAC 

± (2.5 % + 8

 digits

)

 

400 .0AAC 

± (2.8 % + 5

 digits

)

 

 
DC Current 
 

40.00 ADC 

± (2.5 % + 5

 digits

400 .0ADC 

± (2.8 % + 5

 digits

DC Voltage 

400.0 mVDC 

± (0.8% + 2 digits) 

4.000 VDC 

± (1.5% + 2digits) 

40.00 VDC 
400.0 VDC 
600.0 VDC 

± (2 % + 2

 digits

)

 

AC Voltage 
(50-400Hz) 

400.0mVAC 

± (1.5% + 30 digits) 

4.000 VAC 

± (1.5% + 5 digits) 

40.00 VAC 
400.0 VAC 
600.0 VAC 

± (2.0% + 5

 digits

)

 

Resistance 

400.0 

Ω

 

± (1.0% + 4 digits) 

4.000K

Ω

 

± (1.5% + 2 digits) 

40.00K

Ω

 

400.0K

Ω

 

4.000M

Ω

 

± (2.5% + 3 digits) 

40.00M

Ω

 

± (3.5% + 5 digits) 

Capacitance 

40.00nF

 

(4.0% r 20 digits)

 

400.0nF

 

(3% r 5 digits)

 

4.000

μ

40.00

μ

F

 

100.0

μ

F

 

(4.0% r 10 digits)

 

Frequency 

 

10-10kHz 
Sensitivity:100V(<50Hz);50V(50 to 
400Hz;15V(401Hz to 10kHz) 

(1.5% r 2 digits) 

Temp  

(type-K) 

(probe accuracy 
not included) 

-20.0 to 760.0

 (3%rdg+5

C) 

-4.0 to1400.0

 (3%rdg+9

F) 

    

 

 

Clamp size                           

Opening 1.2" (30mm) approx 

Diode Test

 

                   

Test  current of 0.3mA typical; Open circuit voltage 1.5V DC 

typical. 

Continuity Check

 

      Threshold <150

; Test current < 0.5mA 

Low Battery Indication

 

   

“          ” is displayed 

Overrange Indication 

   

“OL””  is displayed 

Measurements Rate          

2 per second, nominal 

Input Impedance

 

       10M

 (VDC and VAC) 

Display 

4000 counts LCD 

AC Current 

50-60Hz (AAC) 

AC Voltage bandwidth  

50-400Hz (VAC) 

Operating Temperature 

41 to 104

o

F (5 to 40

o

C)

 

Storage Temperature

 

  -4 to 140

o

F (-20 to 60

o

C) 

Operating Humidity 

Max 80% up to 87

F (31

C) decreasing linearly to 50% at 

104

F(40

C) 

Storage Humidity 

<80%  

Operating Altitude 

7000ft. (2000meters) maximum. 

Over voltage

 

 

Category III 600V 

Battery

 

One 9V Battery

 

Auto OFF

 

approx. 30 minutes  

Dimensions/Weight

 

     197x70x40mm/183g 

Safety                                    

For indoor use and in accordance with Overvoltage Category 

 

 

 

Ω

1

2

3

4

5

6

7

8

9

10

Отзывы: