ENTERING DATA
3.5
IDENTIFICATION DATA
•
Programmed start.
The oven can be programmed to come on at a set time even when no members of personnel
are in the laboratory ( see paragraph 7.3.1 ).
The time interval is expressed in:
- hours ( 0:99 )
- minutes ( 0:59 ).,
with a resolution of 1 minute.
Use of the delayed start feature is not, of course, necessary for proper oven operation.
•
Entering temperature data.
Temperature, expressed in degrees centigrade (°C) may vary within the following range:
- minimum (room temperature)
- maximum (1,100°C).
•
Entering data for fast temperature increase rate.
The temperature increase rate, expressed in degrees centigrade (°C) per minute may vary
within the following range:
- minimum (1°C per minute)
- maximum (30°C per minute).
•
Entering data for a inter-stage “holding” time.
An inter-stage holding time, lasting anything between 1 minute and 21 hours, can be set.
The equipment described in this manual has an ID platelet giving information about both the
equipment and the manufacturer:
Zhermack S.p.A.
BADIA POLESINE - ITALY
C305315 - DE 1
Matr. 00000000
230 V~ 50/60 Hz 1700W
A
B
C
D
The equipments has been built in compliance with the EC directives listed in the enclosed "EC
declaration of conformity". When requesting spare parts, information or assistance from your
Authorised Servicing Centre always quote the data on the ID plate.
Should the ID plate deteriorate and become illegible, even partially, order a replacement plate
from the manufacturer: remember to quote the relevant data.
A.
Name and address of manufacturer
B.
Equipment code and model
C.
Serial number
D.
Power supply information
Do not remove and/or damage the plate as you must be able to read the data at